Art: 714377-00K
Rev. Date: 31-Jan-12
15-3
Acceptable Criteria
Target values (determined by testing multiple ampules of each level using
multiple lots of i-STAT cartridges with analyzers that have passed the Electronic
Simulator test) are printed on a Value Assignment Sheet posted on the APOC
website at www.abbottpointofcare.com.
Calibration throughout the reportable range of each analyte is verified if each
analyte value falls within the corresponding range in the Value Assignment
Sheet.
If the result for a level is outside the range published in the Value Assignment
Sheet, two additional cartridge runs should be performed on this level and the
three results averaged and then compared to the Value Assignment Sheet range.
If this average value is still outside the acceptable range, troubleshooting may
be required.
Note:
If the Calibration Verification Set is to be used to assess linearity, plot
the analyte value against the mean value of the acceptable range. The
concentrations of analytes in the Calibration Verification Set are not
intended or prepared to be equally spaced.
If testing at extreme altitude refer to Correction of
P
O
2
at Extreme Altitude
under Controls for Blood Gas/Electrolyte/Metabolite Cartridges in the Quality
Control section of the manual.
Overview of Procedure
i-STAT recommends that each sensor type be included in the Calibration
Verification procedure using a selection of analyzers that have passed the
Electronic Simulator check. See the Technical Bulletin "Calibration Verfication
and the i-STAT System" for more information.
Calibration Verification
Solutions for CHEM8+
Cartridges
A five-level Calibration Verification Set is available to verify the calibration of
i-STAT CHEM8+ cartridges throughout the reportable ranges for:
Sodium
TCO
2
Glucose
Potassium
BUN/Urea Creatinine
Chloride
Ionized Calcium
There are four 1.7mL glass ampules of each level in the set.
An additional i-STAT CHEM8+ Calibration Verification Level 1b is available for
purchase to verify the calibration of the i-STAT CHEM8+ TCO
2
at the low end
of the reportable range.
Reactive Ingredients
See the table on page 14-6 of the Quality Control section for full
information.
Storage
Refrigerated storage at 2 to 8 °C (35 to 46°F) should be maintained until
the printed expiration date on the box and ampule labels. i-STAT CHEM8+
Calibration Verification fluids may also be stored at room temperature for up
to 5 days (18 to 30 °C or 64 to 86°F). Prolonged storage at temperatures greater
than 30 °C (86°F) may cause changes in the values of some analytes. Do not
use beyond the expiration date on the box and ampule labels.
Ampule Use
A separate ampule must be used for each cartridge being tested.
i-STAT CHEM8+ CALIBRATION VERIFICATION SET
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......