Art: 714377-00K
Rev. Date: 31-Jan-12
15-5
Calibration throughout the reportable range of each analyte is verified if each
analyte value falls within the corresponding range in the Value Assignment
Sheet.
If the result for a level is outside the range published in the Value Assignment
Sheet, two additional cartridge runs should be performed on this level and the
three results averaged and then compared to the Value Assignment Sheet range.
If this average value is still outside the acceptable range, troubleshooting may
be required.
Note:
If the i-STAT CHEM8+ Calibration Verification Set is to be used to assess
linearity, plot the analyte value against the mean value of the acceptable
range. The concentrations of analytes in the i-STAT CHEM8+ Calibration
Verification Set are not intended or prepared to be equally spaced.
i-STAT CARDIAC MARKER CALIBRATION VERIFICATION SET
A three level Calibration Verification Set is available to verify the calibration of
cTnI cartridges throughout the reportable range.
There are two 1.0 mL plastic vials of each level in the set.
Intended Use
i-STAT
®
Cardiac Markers Calibration Verification Controls are assayed human
sera used to verify the calibration of the cTnI test throughout the reportable
range.
Warnings and
Precautions
Handle this product using the same safety precautions used when handling any
potentially infectious material. The human sera used in the preparation of this
product have been tested by FDA approved test methods and found negative/
non-reactive for HIV-1, HIV-2, HBsAg, HCV, HTLV-1 and HTLV-2. However, no
known test method can offer complete assurance that products derived from
human blood will not transmit infectious disease.
Each vial contains < 0.1% sodium azide as a preservative. Dispose of this product
according to all local, state and national regulations. If disposed down a drain,
the sodium azide in this product may react with lead and copper plumbing to
form highly explosive metal azides. Flush drain with large amounts of water
to prevent azide build-up.
Do not use if the controls arrive thawed or uncapped.
Bacterial contamination can cause an increase in turbidity. Do not use the control
if there is visible evidence of microbial growth or gross contamination.
Storage and Stability
Store at
≤
–18°C (-1°F) in a non-defrosting freezer. After thaw, the opened or
unopened vial is stable for 4 hours when capped and stored at 2-8°C (35-46°F).
Do not re-freeze
Procedure
Step
Action
1
Remove vial from freezer and thaw at room temperature (18-30°C) for 15 minutes.
2
Thoroughly mix by gently swirling the bottle. Avoid foaming of the sample.
3
Dispense a drop of sample directly from vial into the i-STAT cTnI cartridge and seal the
cartridge. Tightly recap the bottle immediately after all sampling is complete and store
at 2-8°C.
4
Insert cartridge into an i-STAT analyzer.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......