PROCEDURE MANUAL FOR THE i-STAT SYSTEM
2
REV. DATE: 16-Oct-12
ART: 714446-00L
Controls
i-STAT Controls for blood gases, electrolytes, and chemistries
Store at 2 to 8°C (35° to 46°F). Controls may be stored at room temperature (18 to 30°C or 64 to 86°F) for five days. Do not
use after expiration date on the box and ampules.
i-STAT Controls for ACT and PT/INR
Store at 2 to 8°C (35° to 46°F). Do not use after expiration date on the box and vials. Controls should be used immediately
after reconstitution.
i-STAT Controls for cTnI and BNP
Store at
≤
-18°C (-1°F) in a non-defrosting freezer. After thawing, the opened or unopened 1.0 mL vial is stable for 4 hours
when capped and stored at 2 to 8°C (35° to 46°F). Do not refreeze
CLINIQA Liquid Cardiac Marker Control for CK-MB
This control requires no reconstitution or frozen storage. It is stable until the expiration date on the vial label when stored
unopened at 2 to 8°C. Once opened, CLINIQA Liquid Cardiac Marker Control is stable for 30 days when stored tightly
capped at 2 to 8°C.
RNA Medical Hematocrit Control
RNA Medical Hematocrit Control is stable until the expiration date stated on the ampule when stored at temperatures of
2 to 25°C. Do not freeze or expose ampules to temperatures greater than 30°C. If stored refrigerated, the control material
should be equilibrated to room temperature for at least 4 hours prior to testing.
Electronic Simulator
Store at room temperature and protect contact pads from contamination by replacing the plastic cap and placing the
Electronic Simulator in its protective case after use.
BLOOD SPECIMENS
Blood Collection Equipment
Cartridges for Blood Gas/Electrolytes/Chemistries/Hematocrit
Skin puncture: lancet and capillary collection tube (plain, lithium heparin, or balanced heparin for electrolytes and
blood gases)
Venipuncture: lithium or sodium heparin collection tubes and disposable transfer device.
Arterial puncture: Plain syringe or blood gas syringe with heparin and labeled for the assays performed or with the
least amount of heparin to prevent clotting (10 U heparin/mL of blood)
Cartridges for ACT
Skin puncture: not recommended
Venipuncture and arterial puncture: plain plastic syringe without anticoagulant
Cartridges for PT/INR
Skin puncture: lancet only needed. Cartridge can be filled directly from the finger.
Venipuncture: plain plastic syringe without anticoagulant.
Cartridges for Troponin I/ cTnI and CK-MB
Skin puncture: not recommended.
Venipuncture: lithium or sodium heparin collection tubes and disposable transfer device (e.g. 1 cc syringe and a.
Alternately, a plain syringe or plain collection tube and disposable transfer device can be used if the sample is
tested within one minute of patient draw.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......