background image

P

ROFESSIONAL 

PROCEDURE

GUIDE

MULTI-TEST A1C SYSTEM

Preparation

1

2

3

4

5

7

9

8

6

BEFORE YOU BEGIN

COLLECT BLOOD

INSERT BLOOD 

COLLECTOR

SHAKE

INSERT CARTRIDGE

DISPENSE SAMPLE 

INTO CARTRIDGE

REUSE MONITOR

5 MINUTES TO RESULTS

PREPARE SAMPLER

Blood Collection

Blood Dilution

Blood Testing

Results

°C

°F

• 

Run the test with all 
parts of the test kit at 
the same temperature 
within the specified 
range.

• 

If the kit has recently been at high 
temperatures (above 82ºF) or in the 
refrigerator, keep the kit at room 
temperature for at least one hour before 
use.

• 

Avoid running the test in direct sunlight, 
on hot or cold surfaces, or near sources of 
heat or cold.

• 

Quality control materials should be used 
to confirm the test kit is working properly. 
Refer to the product insert for information 
on when to run controls.

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

LOT 0628216

EXP 2014 01 27

0628216

Lot Number and Dating Label

Monitor (back)

Pouched

Test Cartridge

Ensure Lot #’s match

Complete

test within

15 minutes.

Sampler

Body

Blood

Collector

Base

(do not remove!)

WAIT T

O OPE

N

Or

45º

x

x

x

MATCH 

CODES

SAMPLE

SAMPLE

SAMPLE

A1

A1

% A1C

A09

LOT 0628216

EXP 21 SEP 14

LOT 0628216

EXP 21 SEP 14

LOT 0628216

EXP 21 SEP 14

MA

TCH 

CODES

% A1C

% A1C

MA

TCH 

CODES

SAMPLE

SAMPLE

SAMPLE

A09

A09

% A1C

% A1C

SAMPLE

SAMPLE

SAMPLE

A09

% A1C

SAMPLE

SAMPLE

SAMPLE

A09

A09

% A1C

MATCH 

CODES

Venous Draw

Mix blood 
well before 
testing

Too little 

add more 
blood

Not fully inserted 
Keep pushing!

* If not, call Customer Service at 
1-877-870-5610.

If you cannot resolve an error, please 
call Customer Service at  
1-877-870-5610.

Do not handle 
Monitor again until 
test is complete!

Do Codes 
Match? *

Monitor and Test Cartridge codes 
must match

Ensure Monitor is on level surface

WAIT 

for SMPL to display

Ready for Sampler

Fully inserted

Just right

Too much 

wipe away 
excess

Collect blood from a slide

Push firmly!

Fingerstick

Shake well 6-8 times 

This will mix the blood 

with the solution 

Push down completely to 
dispense diluted sample 
Remove quickly 

Stand Sampler on table 

while preparing Cartridge

“Click” Test Cartridge into place

Use your own 
lancet device 
to draw blood

Fully insert 
Blood Collector into 
Sampler Body

Keep
pushing!

Twisting 

motion 

helps

Gently touch blood 
drop to fill

OPEN NOW

Use within 2 minutes.

Remove
Base

Display 

counts down

QC test result*

# of tests left

Result

This result cycle remains displayed for 
60 minutes or until the next Test Cartridge 
is inserted.

THE MONITOR IS REUSABLE

To run another test, use a new 
Sampler and Test Cartridge from the 
same kit and return to Step 1, 

PREPARATION.

ALWAYS MATCH LOT NUMBERS

Use Monitor only with the materials 
included in the original kit. The Monitor 
will expire after the programmed number 
of tests have been run. If another Test 
Cartridge is inserted, the Monitor will 
display “00 TL.”

* If “QCOK” is not displayed, please see 
list of error codes on reverse side. 

Discard 
Test Cartridge

Save 
Monitor 

Lot

Code

Product Dating 
& Lot ID Label

Lot
Number

91078 B 3/2014

Reviews: