Code
Number
Cause/Action
Message on Display
Explanation
128, 131,
132, 134,
135 - 138
Cartridge Error /
Use
Another Cartridge
These codes are most often related to poor filling of an
immunoassay cartridge, the presence of sample bubbles, or the
abrupt insertion of a cartridge into the analyzer.
Guidelines for proper filling:
1.
Discard
(always)
1 drop from delivery device to clear
unseen bubbles.
2.
Hang
single drop slightly larger than round target well.
3.
Touch
one drop
(only)
to round target well allowing
cartridge to draw sample in.
4.
Confirm
sample volume lines up with top of RED FILL
LINE diagram.
5.
Close
slide cover from left to right.
Guidelines for cartridge insertion:
1. After closing the cartridge, grasp the cartridge closure
between your first finger and thumb. There is a recess
for your thumb in the closure.
2. Guide the cartridge into the analyzer gently, until a soft
click is heard.
129, 142,
143
Cartridge Error /
Use
Another Cartridge
The analyzer detected analysis fluid mixed with the sample. Try
another cartridge.
130
Cartridge Error /
Use
Another Cartridge
The analyzer detected an air bubble in the sample segment. Try
another cartridge.
140
Lot Expired
The analyzer detected an expired cartridge lot. Check the
expiration date and repeat the test using a non-expired cartridge
lot.
141
Test Canceled by
Operator
This code will be displayed if the cartridge barcode is not
scanned within 60 seconds of cartridge insertion. The correct
barcode to scan is the barcode on the cartridge portion pack, not
the one on the cartridge box. An example of the portion pack
barcode is found in the table listing for code 69 above.
145
Cartridge Error /
Use
Another Cartridge
The analyzer failed to detect fluid arrival upon the initial sample
push. This may be caused by a(n):
•
cartridge leak,
•
failure to close the cartridge completely. Ensure that the
slide cover is fully engaged before inserting the cartridge
into the analyzer
•
underfilled cartridge. Once a single drop of sample is
touched to the target well, immunoassay cartridges
will fill automatically by wicking the sample at a fixed
speed. Trying to inject the sample into the cartridge or
adding more sample to the target well will not make the
cartridge fill faster. Wait for the sample to reach the “fill
to” mark, and then close the cartridge.
146
Cartridge Error /
Use
Another Cartridge
Overfilled cartridge. Repeat the test.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......