Art: 723270-00A
Rev. Date: 20-Apr-12
ampule to room temperature for 4 hours prior to use. If
P
O
2
is not being measured,
equilibrate the ampule for approximately 30 minutes at room temperature.
Procedure
STEP
ACTION
1
Access the Cal Ver option under Quality Tests in the
Administration Menu. Enter the required information. The
handheld allows 15 minutes (or the customized timeout) to
insert the cartridge after the last data entry.
2
Immediately before use, shake the ampule vigorously for 5 to
10 seconds to equilibrate the liquid and gas phases.
To shake, hold the ampule at the tip and bottom with forefinger
and thumb to minimize increasing the temperature of the
solution. If necessary, tap the tip of the ampule to send
solution back into the bottom section of the ampule.
3
Protect fingers with gauze, tissue or glove, or use an ampule
breaker to snap off the tip of the ampule at the neck.
4
Immediately transfer the solution from the ampule into a
capillary tube or syringe, and then immediately transfer the
solution into a cartridge.
5
Immediately seal the cartridge and insert it into a handheld – it
is important not to expose the solution to room air since this
will alter the results.
Note:
Since aqueous based solutions such as control
materials lack the buffering capability of whole blood,
the transfer process from ampule to cartridge must be
more expedient than with a patient sample.
Transfer with Capillary Tube
Plain capillary tubes are recommended to transfer an aqueous calibration verification
solution from the ampule to the cartridge. When using a capillary tube (fresh capillary
tubes with sufficient fill capacity are recommended), fill from the bottom of the ampule
to avoid drawing air into the capillary tube. Avoid drawing solution from the surface by
placing a finger over the far end of the tube as it is inserted into the ampule. Once the
open end of the tube rests at the bottom of the ampule, uncover the other end to allow
filling by capillary action.
Transfer with Syringe
Plain syringes (fresh 1cc or 3cc sterile syringe with 16 – 20 gauge needles) are
recommended to transfer aqueous calibration verification solutions from the ampule to
the cartridge. When using a syringe, slowly draw approximately 1mL of solution from
the bottom of the ampule.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......