1
Press
Login
and
scan or enter
Operator ID.
Press
Accept.
6
Enter sample to the end of the strip by
touching the strip to the QC drop or patient
blood. Keep the meter on flat surface or upright.
Gently mix the control solution before use and
discard the first drop of control material.
Note:
Do not get sample material into the strip port.
3
Scan strip lot
number from
the barcode at the
test strip vial and
press
Accept.
7
The meter will display the result. Patient
test result can be accepted my pressing
the
Accept
button. By pressing the
Reject
button
result can be rejected and a comment can be
added to the result with the
Comment
button.
For QC material the meter will show
Pass
or
Fail
to indicate if QC test was passed.
Use ejector button on the back of the meter to
eject the used strip.
2
Switch to quality
control test by
pressing
QC
or press
Accept
to perform
a patient test.
4
Scan or enter
patient ID or QC
material LOT number.
Press
Accept.
5
Insert test strip
into meter.
www.aidian.eu
Accept
Accept
Clear
Clear
Accept
Review
Manual
QC
MENU
Logout
Glu
Patient Test
NICU-1
Batt:
Mem:
Review
Login
Press OK to begin testing
Location:
NICU
Facility:
Your Facility
View
Setup
Welcome
Welcome
Apply Sample
QC Result
Patient Result
Patient Test
Enter Strip Lot
Enter QC Lot
Insert Strip
Insert Test Strip in meter
Cancel
Touch test strip to Level 1 QC drop
Remove test strip to cancel test
Remove test strip to cancel test
Accept
Comment
Accept
Reject
Comment
QC Result
Op: 8636
QC Lot: 0213806081
Lev
el:
1
QC
Pending
07/19 06:19
Strip Lot: 0206806099
Glu
4.4
mmol/L
PASS
Range 4.0-4.8
Patient Result
Pt ID: 194546646
Darryl Smith
Strip Lot: 0206806099
5.7
mmol/L
Glu
Normal 3.9-6.1
07/19 06:19
Patient
Pending
Please consult
the instructions
for use before
performing
the test.
Additional
features may be
enabled for the
meter.
Brief instructions
Nova StatStrip Glucose and
ß-Ketone Hospital Meter
Touch test strip to blood drop