![background image](http://html1.mh-extra.com/html/abbott/i-stat-1/i-stat-1_system-manual_450958437.webp)
Rev. Date: 01-Aug-11
Art: 714178-00L
HB - 3
Method Comparison (%PCV)
Coulter
®
S Plus
Nova STAT Profile
®
5
Abbott
Cell-Dyn 4000
Sysmex SE9500
n
142
192
29
29
Sxx
0.50
0.46
0.41
0.53
Syy
1.09
1.31
0.77
0.76
Slope
0.98
1.06
1.06
1.11
Int’t
1.78
-3.98
-1.42
-4.19
Sy.x
2.03
2.063
1.13
0.98
Xmin
18
21
19
24
Xmax
51
50
46
47
r
0.952
0.932
0.993
0.980
Factors Affecting Results*
The measurement of certain blood samples with high erythrocyte sedimentation rates (ESR) may be
affected by analyzer angle. While testing blood samples, beginning ninety (90) seconds after the cartridge
is inserted, the analyzer should remain level until a result is obtained. A level surface includes running the
handheld in the downloader/recharger.
Interferent Effect
WBC
Grossly elevated white blood cell counts may increase results.
Total Protein
Hematocrit results are affected by the level of total protein as follows:
Displayed Result
TP < 6.5 g/dL
TP > 8.0 g/dL
HCT < 40 %PCV
Hct decreased by ~1% PCV
for each decrease of
1 g/dL TP
Hct increased by ~1% PCV
for each increase
1 g/dL TP
HCT > 40 % PCV
Hct decreased by ~0.75 % PCV
for each decrease of
1 g/dL TP
Hct increased by ~0.75 %PCV
for each increase
1 g/dL TP
Total protein levels may be low in neonatal and burn patient populations, as well
as in additional clinical populations listed in Statland
3
. Total protein levels may also
be decreased in patients undergoing cardiopulmonary bypass (CPB) or ECMO,
and with patients receiving large volumes of saline-based IV fluids. Care should be
taken when using hematocrit results from patients with total protein levels below
the adult reference range (6.5 to 8 g/dL).
The CPB sample type can be used to correct the hematocrit result for the dilutional
affect of the pump prime in cardiovascular surgery. The CPB algorithm assumes
that cells and plasma are diluted equally and that the pump priming solution has
no added albumin or other colloid or packed red blood cells. Since perfusion
practices vary, it is recommended that each practice verify the use of the CPB
sample type and the length of time in which the CPB sample type should be used
during the recovery period. Note that for hematocrit values above 30 %PCV, the
CPB correction is
≤
1.5 %PCV; the size of the correction at this level should not
impact transfusion decisions.
Lipids
Abnormally high lipids may increase results. Interference from lipids will be about
two thirds the size of the interference from protein.
Sodium
The sample electrolyte concentration is used to correct the measured conductivity
prior to reporting hematocrit results. Factors that affect sodium will therefore also
affect hematocrit.
*It is possible that other interfering substances may be encountered. These results are representative and your results may differ somewhat due to
test-to-test variation. The degree of interference at concentrations other than those listed might not be predictable.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......