Celite ACT - 7
The analyzer should remain on a level surface with the display facing up during testing. If the analyzer is not
level, the ACT result may be affected by more than 10%. A level surface includes running the handheld in
the downloader/recharger.
Hemodilution may affect test results.
Platelet dysfunction, hereditary or acquired, may affect the results of this test. This includes the
administration of pharmacological compounds known as platelet inhibitors which affect platelet function.
Factor deficiencies, dysprothrombinemias, other coagulopathies, and other pharmacological compounds
may also affect the results of this test.
The i-STAT ACT test is not affected by hematocrit in the range of 20 - 70%, fibrinogen concentration in the
range from 100 - 500 mg/dL, or sample temperature from 15 - 37ºC.
Specimen Collection and Preparation
The i-STAT
Celite
ACT test can be performed using venous or arterial samples.
Venipunctures and Arterial Punctures
• Collection technique resulting in good blood flow must be used.
• The sample for testing should be drawn into a
plastic collection device (either a plastic
syringe or plastic evacuated tube).
• The collection device
cannot contain anticoagulants such as heparin, EDTA, oxalate, or citrate.
• The collection device cannot contain clot activators or serum separators.
• The sample should be immediately dispensed into the sample well of a cartridge.
• If a second measurement is required, a fresh sample must be obtained.
Note: Some experts recommend drawing and discarding a sample of at least 1 mL prior to drawing
sample for coagulation testing.
5
Indwelling line
• Fluid drip through the line must be discontinued.
•
If blood must be drawn from an indwelling line, possible heparin contamination and specimen dilution
should be considered. The line should be flushed with 5 mL of saline and the first 5 mL of blood or
six dead space volumes should be discarded.
• Withdraw the sample for testing into a fresh
plastic syringe.
• The collection syringe
cannot contain anticoagulants such as heparin, EDTA, oxalate, or citrate.
• The sample should be immediately dispensed into the sample well of a cartridge.
• If a second measurement is needed, draw a fresh sample.
Extracorporeal line
• Flush the extracorporeal blood access line by withdrawing 5 mL of blood into a syringe and discard the
syringe.
• Withdraw the sample for testing into a fresh
plastic syringe.
• The collection syringe
cannot contain anticoagulants such as heparin, EDTA, oxalate, or citrate.
• The sample should be immediately dispensed into the sample well of a cartridge.
• If a second measurement is needed, draw a fresh sample.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......