CHECKING THE THERMAL PROBES IN THE i-STAT ANALYZERS
Verification for
Handheld Analyzers
Verify the thermal probe check for the i-STAT Portable Clinical Analyzer and
i-STAT 1 Analyzer as follows:
• External Electronic Simulator used routinely and results transmitted
to a Central Data Station Version 4: On the CDS, click on
System
,
then on
Simulator Results
. Look under the Delta column. Check
that there is a value equal to or less than 0.1 (
≤
0.1) listed for each
analyzer in use in the last 30 days. A value of “--.--” indicates that
the conditions to complete the thermal probe check were not met.
Use the procedure below to check any analyzer that does not have a
numeric value listed.
• External Electronic Simulator used routinely and results transmitted
to a Central Data Station Version 5: On the CDS, click on
Data
Viewer
, then on
Simulator
. Look under the Probe Delta column.
Check that there is a value equal to or less than 0.1 (
≤
0.1) listed for
each analyzer in use in the last 30 days. A value of “--.--” indicates
that the conditions to complete the thermal probe check were not
met. Use the procedure below to check any analyzer that does not
have a numeric value listed.
• External Electronic Simulator used routinely, results not transmitted
to a Central Data Station: Use the procedure below to check the
thermal probes on each analyzer every six months.
• Internal Electronic Simulator used routinely: Use the procedure
below to check the thermal probes on each analyzer twice a year.
Procedure for
Handheld Analyzers
Check the thermal probes on the i-STAT Portable Clinical Analyzer and i-STAT 1
Analyzer as follows:
1. If the analyzer and simulator have been stored separately in areas
where the ambient temperature differs by more than 3°C (5°F), allow
the simulator and analyzer to stand in the same place, out of drafts,
for 30 minutes before inserting the simulator into the analyzer.
Handle the simulator as little as possible to maintain its thermal
uniformity and stability.
2. Insert the simulator into the analyzer.
3. When results are displayed, the difference between the thermal
probes can be viewed on the analyzer’s screen:
• Portable Clinical Analyzer: while holding down the DIS key, press
the 1 key.
•
i-STAT 1 Analyzer: press the period key.
4. Interpretation of the thermal probe check value:
• Acceptable: a value equal to or less than 0.1 (
≤
0.1)
• Not acceptable: a FAIL message with a “t” Quality Check Code or
a value greater than 0.1. Repeat the procedure to confirm results.
Contact your Technical Support representative if the repeat thermal
check value is greater than 0.1.
• Repeat the procedure: if “--.--” is displayed. Take care to handle
the simulator a little as possible. It may help to partially insert the
simulator into the analyzer and let it stand for 15 minutes before
inserting all the way.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......