TEST CYCLE MESSAGES AND QUALITY CHECK CODES
Overview
If a problem is detected during a testing cycle, the cycle will be stopped and
a message will identify the problem and indicate the next step to be taken. If
the problem causes testing to be disabled, the problem must be corrected and
the analyzer must be turned off and back on before testing will be enabled.
Environmental
Conditions
The following messages usually indicate a condition related to the environment
or the state of the analyzer. These conditions are usually benign and go away
after the the offending condition is corrected.
Message on Display
Cause
Action
Date Invalid, Check Clock
The analyzer will not allow a
date that precedes or exceeds
the six months lifetime of the
CLEW software.
Press
Menu
once to go to the Test
Menu and then again to go to the
Administration Menu. Press 5
to go to the Set Clock screen and
correct the date.
Dead Batteries, Replace
Batteries
There is insufficient battery
power to complete a test cycle.
Change the disposable lithium
batteries or recharge the
rechargeable battery.
Temperature Out of Range,
Check Status Page
The analyzer makes a
temperature measurement
before initiating a test cycle.
Check the temperature reading
on the Analyzer Status screen
(under the Administration Menu).
If below the operating range,
move to a warmer area. If above
the operating range, move to a
cooler area. Allow time for the
analyzer to equilibrate to the new
temperature. Check the Analyzer
Status screen periodically.
Invalid or Expired CLEW
The software has become
corrupt or has expired. The
Product Update for each
software update includes the
expiration date.
Verify that the date in the analyzer
is correct. Change the software
if expired. Update the software
again if not expired. If the
message is displayed again, refer
to Support Services information at
the end of this section.
Analyzer Interrupted, Use
Another Cartridge
The analyzer detected that
the last cartridge run was not
completed. This can happen
if battery voltage is low, or
if batteries were removed or
making poor contact while
a cartridge was still in the
analyzer.
Check that the battery pack
is inserted properly. Turn the
analyzer on and check for the
Low Battery message; replace or
recharge if needed.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......