It is especially important not to expose aqueous samples for blood gases and ionized calcium to air. The
instructions in the Quality Control section of the i-STAT System Manual under “Transfer with Capillary Tube”
or “Transfer with Syringe” should be followed.
• Take care not to draw air bubbles into the transfer device. If there is only one ampule per level, the
opened ampule may be exposed to air for too long to fill a second transfer device. Exposure to air
could affect pH,
P
CO2,
P
O2 and ionized calcium results.
• If air is trapped between the leading edge of the solution and the syringe plunger, do not invert
the syringe to expel it. Air trapped near the plunger will not affect results from the solution near the
tip of the syringe, but may affect the sample as it passes through the syringe when being expelled.
• One or two drops should be expelled from a syringe before filling a cartridge.
PRofICIenCy TeST fAILuReS
Certain PT survey samples are not compatible with i-STAT System sensors and may contain interfering
substances that could affect results. Fluorocarbon samples for blood gases and fixed-cell samples for
hematocrit are not compatible with the sensors.
Occasionally, results from non-whole blood samples may differ between older and newer generations
of cartridges. These differences may be significant. Abbott Point of Care will work with PT providers
to prevent i-STAT System users from being unfairly penalized for PT failures that can be attributed to
manufacturing changes. This is why it is important to record the cartridge type and lot numbers used
to test samples. Note: the CLEW software prevents manufacturing changes from affecting results when
testing patient samples.
PRofICIenCy TeSTIng SuRvey PRovIdeRS
Note 1: This list may not include all providers and is not intended as an endorsement of any particular
provider. Abbott Point of Care no longer evaluates surveys and assumes that for those surveys
that list the i-STAT System, compatibility has been confirmed by third-party testing or a history
of adequate results.
Note 2:
Select a survey for hematocrit which is formulated for a conductometric method
Proficiency Provider
Survey Title(s)
Additional Information or
Recommendations
College of American Pathologists (CAP)
College of American
Pathologists (CAP)
325 Waukegan road
Northfield, IL
60093-2750
800-323-4040 or 847-832-7000
www.cap.org
CAP AQ3 survey for blood gases and
electrolytes: includes iCa, Cl, Hct, Hb,
Lac,
P
CO
2
, pH,
P
O
2
, K, Na, and TCO
2
CAP AQ4 survey for blood gases and
electrolytes: includes all AQ3 analytes
plus Glu, BUN, and Crea.
CAP Excel L24: includes iCa, Cl, Hct,
P
CO
2
, pH,
P
O
2
, K, and Na.
AQ3 or AQ4 is recommended if L24
does not cover all analytes
P
O2 results should not be
corrected for barometric pressure.
The barometric pressure should
be reported with the uncorrected
results.
Barometric pressure can
be reported from the Status Page
after the results are displayed.
Note:
Use the
green labeled
samples for hematocrit testing.
CAP CT1 survey for Activated Clotting
Time (ACT)
Samples are reconstituted with
0.025 mol/L (25 mmol/L ) calcium
chloride
supplied by the user.
CAP Plasma Cardiac Markers (PCArM)
Survey
XL/XLN Excel Series L30
For use with i-STAT cTnI, CK-MB,
and BNP cartridges.
Whole Blood PT/INr (WP3) and Excel
PT/INr XC19
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......