7 - 15
Art: 714369-00H
Rev. Date: 03-Aug-12
•
Do not operate the printer without paper.
•
Do not allow the power supply to become a trip hazard.
•
Do not disturb the handheld or printer until printing is complete since this will interrupt the printout. If
printing is interrupted, realign the printer and handheld or replace the handheld in the Downloader to
resume printing. Note: if significant time has elapsed, some results may be missing from the printout.
Reprint the results.
•
If printed results appear inconsistent with a patient’s clinical assessment, verify that the printed results
match the data in the handheld. If the results do match, the patient sample should be retested using
another cartridge. If they do not match, reprint the results. If the reprint still does not match the
handheld data, the printer requires service and the printed results must not be used.
•
Skin irritation, including caustic burns/injury, may occur following exposure to a leaking battery.
Always wear gloves when handling a leaking battery, and do not permit a leaking battery to contact
skin. Should skin exposure to a leaking battery occur, follow the first aid measures outlined in the
MSDS Sheet for the Novacell nickel metal hydride battery.
PRINTER TROUBLESHOOTING
PRINTER SYMPTOM
RECOMMMENDED ACTION(S)
Printer is not printing. The POWER indicator
light is green/orange and the STATUS indicator
light is green.
•
Check that the results are displayed on the
handheld, or that results have been selected
from
List
under
Data Review.
•
If printing directly from the handheld, check
that the distance between the analyzer and
printer is not too short or too long.
•
Perform printer self test to ensure that
printer is functioning. Turn the printer off.
While pressing the Paper Feed button,
press down on the Power button until the
printout begins, and then let go of both
buttons. Ensure that the resulting printout is
clear and complete.
Printer is feeding paper, but nothing is printed.
Check that the paper is feeding from under the
roll.
Printer is not printing and POWER indicator is
red.
Battery needs to be recharged.
Printer POWER indicator does not illuminate
when printer is turned on.
Battery needs to be recharged.
Printer is not printing and STATUS indicator is
orange.
Printer is out of paper.
Printer is not printing and STATUS indicator is
red.
Print head temperature is out of range. Print-
ing will be inhibited until print head temperature
returns to normal level.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......