3-3
Rev. Date: 06-Aug-12
Art: 714365-00G
Standardization and
Calibration
Standardization is the process by which a manufacturer establishes “true”
values for representative samples. The sensors in the i-STAT cartridges are
standardized against plasma methods used by major laboratory systems or,
for blood gases, against tonometry. A multi-point calibration curve, the
slope or sensitivity of which is defined by coefficients in the CLEW software,
is derived for each sensor by this standardization process. These calibration
curves are stable over many lots and only need to be adjusted if a change in a
manufacturing process affects the curve or if the relationship between results
on the i-STAT System and other major laboratory systems drifts. For the
convenience of users, CLEW updates are scheduled two times a year.
A one-point calibration is performed each time a cartridge requiring calibration
is used. During the first part of the testing cycle, the calibrant solution is
automatically released from its foil pack and is positioned over the sensors.
The signals produced by the sensors’ responses to the calibrant solution
are measured. This one-point calibration adjusts the offset of the stored
calibration curve. Next, the analyzer automatically moves the sample over the
sensors and the signals produced by the sensors’ responses to the sample are
measured. While coefficients are used rather than graphic calibration curves,
the calculation of the result is equivalent to reading the sample’s concentration
from adjusted calibration curve.
Packaging
Each cartridge is sealed in a foil pouch or clear plastic portion pack for
protection during storage.
Labeling on the carton, box and pouch/portion pack identify:
•
the panel name.
•
the tests included in the panel.
•
the lot number.
•
the expiration date of the cartridge.
If the pouch/portion pack has been punctured, the cartridge should not be
used.
14
tage jours giorni días dias dagen dage
dagar päivää dager
ημέρες
gün
days
Abbott Point of Care Inc.
Abbott Park, IL 60064, USA
Product of Canada
Room Temperature Storage
Lagerung bei Raumtemperatur • Stockage à température ambiante • Conservazione a temperatura ambiente •
Almacenamiento a temperatura ambiente • Armazenamento da temperature da sala • Opslag bij kamertemper-
atuur • Opbevaring ved stuetemperatur • Förvaring i rumstemperatur • Säilytys huoneenlämmössä • Oppbevar-
ing ved romtemperatur • Φυλάσσεται σε θερμοκρασία δωματίου • Oda Sıcaklığında Saklama
EC8+
Na K Cl Glu pH
P
CO
2
BUN/Urea Hct
HCO
3
T
CO
2
BE AnGap Hb
30ºC
86ºF
18ºC
64ºF
30ºC
86ºF
18ºC
64ºF
8ºC
46ºF
2ºC
35ºF
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......