PROCEDURE MANUAL FOR THE i-STAT SYSTEM
23
REV. DATE: 16-Oct-12
ART: 714446-00L
Target values are specific to the i-STAT System. Results may differ if used with other methods (i.e., other IVD
instrumentation).
Always remember to analyze the control material in the Control pathway and the calibration verification material in the
Cal Ver pathway under the Quality Tests option of the i-STAT 1 Analyzer Administration Menu.
Remedial Action:
If any results are outside the published expected ranges:
DO NOT USE cartridges from the suspect lot.
Quarantine the suspect lot.
Notify the i-STAT System Coordinator immediately.
Record the QC failure in the i-STAT QC Action Log along with the action taken.
Procedures for for testing CK-MB cartridges with CLINIQA Cardiac Marker controls for i-STAT
1.
CLINIQA Liquid QC Cardiac Marker Control is a ready-to-use liquid control requiring no reconstitution or frozen
storage. It is stable until the expiration date on the vial label when stored unopened at 2-8°C. Once opened,
CLINIQA Liquid QC Cardiac Marker Control is stable for 30 days when stored tightly capped at 2-8°C.
2.
Immediately before use, gently mix the contents of the control vial to ensure homogeneity. Avoid foaming of the
sample.
3.
Open the vial and transfer a drop of the solution into the i-STAT CK-MB cartridge using a plain capillary tube,
plain syringe, or plastic transfer pipette. Tightly recap the control vial and store it at 2-8°C.
4.
Seal the cartridge and immediately insert it into the i-STAT 1 Analyzer.
5.
Compare the i-STAT CK-MB result to the Value Assignment Sheet value. Always ensure that the lot number and
software revision on the Value Assignment Sheet matches the lot number of the vial in use and the software
revision in the handheld. Should results fall outside the range, refer to the Troubleshooting section of i-STAT 1
System Manual section 14 (Quality Control).
6.
If available, transmit results to the Central Data Station.
Remedial Action
If any results are outside the published expected ranges:
DO NOT USE cartridges from the suspect lot.
Quarantine the suspect lot.
Notify the i-STAT System Coordinator immediately.
Record the QC failure in the i-STAT QC Action Log along with the action taken
.
Procedure for testing BNP cartridges with i-STAT BNP Level 1, 2, or 3 Controls
1.
Remove vial from freezer and thaw at room temperature (18-30°) for 15 minutes.
2.
Gently invert the vial 10 times, then swirl the vial 10 times. Inspect the sides of the vial to ensure that no particulate
matter is clinging to the sides of the vial. If solids are observed in the control fluid or on the vial wall, repeat the
mixing procedure. If further mixing does not homogenize the sample, discard the vial and thaw a fresh vial.
3.
Open the vial and transfer a drop of solution directly into the BNP cartridge and seal the cartridge. If short term
storage (<4 hours) is desired, tightly recap the bottle immediately, and store at 2-8°C.
4.
Seal the cartridge and insert into the i-STAT 1 Analyzer.
5.
Compare the i-STAT BNP result to the Value Assignment Sheet ranges. Always ensure that the lot number and the
software revision on the Value Assignment Sheet matches the lot number of the vial in use and the software
revision in the handheld. Should results fall outside the range, refer to the Troubleshooting section of i-STAT 1
System Manual section 14 (Quality Control).
6.
If available, transmit results to the Central Data Station.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......