Self-Monitoring Blood Glucose System

Please read this manual thoroughly before first 
using this device

Summary of Contents for Sprint nG

Page 1: ...ENGLISH User Manual HMM Diagnostics GmbH Friedrichstra e 89 D 69221 Dossenheim Germany www hmm info Self Monitoring Blood Glucose System Please read this manual thoroughly before first using this devi...

Page 2: ...44...

Page 3: ...he smartLAB nG test strips 54 Explanation of symbols 55 Set Content 56 III Setup functions 57 Insert and changing batteries 57 Setting Time and Date 59 Using smartLAB nG Blood Glucose Test Strips 60 F...

Page 4: ...46 IV Miscellaneous 74 Maintenance 74 Frequently Asked Questions 74 Troubleshooting 77 Limitations of the measurement procedure 80 Regularity and Safety Notice 83 Warranty 83...

Page 5: ...vides you and your diabetes care team with important information and step by step direction to use the system correctly Although your smartLAB sprint nG system is easy to use you should con sult your...

Page 6: ...uce gluconolactone During the reaction a mediator transfers electrons to the electrode surface and generates a current The amount of the current is pro portional to the amount of glucose present in th...

Page 7: ...results Severe impact may cause the meter to malfunction Do not dis assemble the meter as it may damage components inside and cause an incorrect reading The warranty will be void if the meter has bee...

Page 8: ...strips immediately after 90 days from open date The strips are for single use only You should not test critically ill patients with home use blood glucose meters Warning for potential biohazard Health...

Page 9: ...molar hyperglycemic nonke totic coma HHNKC you may get a test result that is lower than what your blood glucose really is If you think you are dehydrated call your healthcare professional right away I...

Page 10: ...e noon PM Indicate after noon Batteries status When the battery symbol appears prepare new batteries for installation AC ante cibos Indicate before meal PC post cibos Indicate after meal Insert strip...

Page 11: ...ct 30 55 12 Battery type 2 1 5 V AAA Lithium batteries 13 Battery life over 2000 measurements 14 Display size 43 x 35 mm 15 Meter dimensions 98 x 57 x 22 mm 13 Weight 53 2 g without batteries Test Str...

Page 12: ...est strips Note You can find the test strip expira tion date on the vial label next to the symbol Expiration date Vial opening date please note Electrode Insert direction Grip zone Reaction zone Apert...

Page 13: ...y Single use only Storage temperature Manufacturer Before use read manual first Catalogue number Serial number Control solution Opening date of test strip vial Sufficient for Blood glucose test result...

Page 14: ...r 1 User manual 1 carrying bag 2 AAA 1 5V Alkali batteries 1 smartLAB lancing device 1 smartLAB Check Strip 1 smartLAB nGQuick Start instruction 10 smartLAB lancets 10 smartLAB nGblood glucose test st...

Page 15: ...ies are replaced Meter batteries lancets test strips etc must be disposed of according to local regulations at the end of their usage 1 Open the battery door on the back of the meter by pushing the ta...

Page 16: ...er will not turn on the first time batteries are inserted Please press and hold right button or insert the test strip to turn your meter on The meter will turn off automatically Or you can press and h...

Page 17: ...means the time from 12 00 o clock until 24 00 o clock 1 Press the right button to turn the meter on 2 The display shows the last 2 digit of the year that flashes at the top of display as well Press t...

Page 18: ...t out of the bottle The strip is for single use only Do not reuse it Record the date you open the test strip bottle Be sure to check the expiration date on the test strip bottle The test strip is good...

Page 19: ...check strip is inserted with writing to top The meter will turn on automatically and the display will show CHK 2 Check message in LCD display The function check will take approx 3 seconds The check ca...

Page 20: ...ower or higher than expected You are practicing the test procedure Note Professional users are instructed to follow federal state and local guidelines concerning QC practices About the smartLAB nG con...

Page 21: ...would like to purchase smartLAB nG control solutions please contact your local dealer Running a smartLAB nG control solution test Please make sure you have all necessary items at hand when run ning a...

Page 22: ...strip 8 The meter starts to count down from 5 seconds and then shows the test result 9 Do not remove the test strip yet Check if the rea ding falls within the range printed on the test strip bottle 10...

Page 23: ...p vial capped tightly Or was test strip cap left open If the cap was not tight or the bottle was left uncapped open a new bottle of test strips Do not reuse the strips from the affected bottle Does th...

Page 24: ...hands thoroughly with warm water before collecting the blood sample Rinse your hands thoroughly 2 Collecting a blood sample When the blood drop flashes on the display gently massage the part of your f...

Page 25: ...gainst the side of your finger Press the trigger button Then remove the lancing device from your finger 5 Gently squeeze and or massage your fingertip until a round drop of blood on your fingertip app...

Page 26: ...drop appears on the display Do not reuse lancets Dispose lancets and test strips according to your local regulations 4 Discarding used test strips and lancets Write down the measured result in your pe...

Page 27: ...owing steps Gently massage the desired punctuation site on your arm or hand for a few seconds This will improve the blood circulation Push the lancing device with the transparent cap against the de si...

Page 28: ...mg dL 6 7 mmol L Expected results for diabetic adults The normal fasting blood glucose range for an adult with diabetes is 70 130 mg dL 3 9 7 2 mmol L Two hours after meals the blood glucose range for...

Page 29: ...lts A relative humidity greater than 90 may cause inaccurate results 2 A red blood cell count Hematocrit below 30 may cause higher results Hematocrit above 55 may cause lower results 3 Some studies ha...

Page 30: ...r eating taking medication or exercising If you test yourself in the morning then go to the doctor s office for a blood glucose test The results will probably not match even if you are fasting This is...

Page 31: ...is not lost when you replace the battery You do need to check that the time and date are still correct for future readings See Chapter Setting the time and date 3 Once 480 results are in memory addin...

Page 32: ...o not get any moisture in the test strip slot Please do not spray any cleaning solution directly onto the meter Do not pour liquid into the meter or put the meter under water or liquid To clean the la...

Page 33: ...st strip correctly see 2 Does the display read ERR The inserted test strip was already used Please use a new test strip 5 Blood glucose result too high low Please check Was the test strip exposed to a...

Page 34: ...onal 7 Patients undergoing oxygen therapy may have inaccurate results 8 Altitude up to 3050 meters above sea level has no effect on read ings 9 Test results may be false if the patient is severely deh...

Page 35: ...ur local supplier Battery empty Display LP Battery symbol Solution Replace with new batteries Test strip already used or wet Display Err Insert Solution Use a new test strip System error Display 001 E...

Page 36: ...1 1 mmol L Display LO Solution Test again If the result is still the same try a control solution test and if the con trol solution test falls within the correct range please call your healthcare prof...

Page 37: ...e test continues Solution Relocate the meter to a location with temperature between 10 C 40 C 50 F 104 F Temperature too high Display Ht Thermometer symbol The operating temperature is too high above...

Page 38: ...rips The test strips are intended for single use only 7 DO NOT use iodoacetic acid fluoride or sodium fluoride oxalate as a preservative for blood specimens 8 Hematocrit Test strip results are not sig...

Page 39: ...me use blood glucose meter 14 Elevated cholesterol and triglyceride levels may interfere with the way light is reflected producing erroneous meter results 15 Recent studies have shown that EMI can cau...

Page 40: ...42 0 84 Tetracycline 1 5 mg dL 0 03 mmol L 5 32 3 81 3 20 Tolbutamide 100 mg dL 3 70 mmol L 2 60 12 30 0 89 Bilirubin unconjugated 2 4 mg dL 0 04 mmol L 2 52 4 05 0 23 Uric acid 8 mg dL 0 48 mmol L 2...

Page 41: ...product Please register online under following website www hmm info en registration Wear parts batteries etc are excluded from warranty The WEEE logo on the product and the packaging indicates that it...

Page 42: ...agnostics GmbH Friedrichstr 89 D 69221 Dossenheim Germany mail info hmm info www hmm info Manufacturer Weitere Informationen zu den smartLAB Produkten More information on our smartLAB products www sma...