Power/Set Button
2
Display
3
Right Arrow and Left Arrow Buttons
4
Test Strip Slot
5
Battery Door
6
Infrared (IR) Window
Meter Overview
Start Here
Quick Start Guide
If any of the error messages below appear on the display at any time,
refer immediately to the User’s Manual, Chapter 5, Display and Error
Messages.
E-1
The test strip may be damaged or not properly inserted. Remove and
reinsert the test strip, or replace it if damaged.
E-3
Your blood glucose may be extremely high or a meter or a test strip
error has occurred. If your test result matches how you feel, contact your
healthcare professional immediately. If your test result does not match
how you feel, repeat the blood glucose test and see the User’s Manual,
Chapter 2, Unusual Blood Glucose Results. If the second test result does
not match how you feel, perform a control test with the control solution
and a new test strip. If the control result is within the acceptable range,
review the proper testing procedure and repeat the blood glucose test
with a new test strip. If the E-3 code still appears for your blood glucose
test, your blood glucose result may be extremely high and above the
system’s reading range.
Contact your healthcare professional
immediately.
If the control result is not within the acceptable range, see
the User’s Manual, Chapter 4, Understanding Out-of-Range Control
Results.
E-4
Not enough blood or control solution was drawn into the test strip for
measurement or was applied after the test had started. Discard the test
strip and repeat the blood glucose or control test.
E-6
Blood or control solution was applied to the test strip before the
flashing drop symbol appeared on the display. Discard the test strip and
repeat the blood glucose or control test.
E-7
An electronic error occurred, or in rare cases, a used test strip was
removed and reinserted. Turn the meter off and on, or take the battery
out for 20 seconds and reinsert it. Perform a blood glucose or control
test.
E-8
The temperature is above or below the proper range for the system.
Refer to the test strip package insert for system operating conditions.
Move to an area with the appropriate conditions, wait 5 minutes, and
repeat the blood glucose or control test. Do not artificially heat or cool
the meter.
E-9
The battery is almost out of power. Change the battery now. If the
message reappears after the battery has been replaced, remove the
battery again, press any meter button, then reinsert the battery.
E-10
The time and date settings may be incorrect. Make sure the time
and date are correct and adjust, if necessary.
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim, Germany
www.accu-chek.com
Familiarize yourself with the
Accu-Chek Softclix lancing
device.
1
Get a lancet from the pocket in the
carry case.
2
Remove the cap by pulling it
straight off. Do not twist the cap.
Insert the lancet into the lancing
device until the lancet clicks.
4
3
5
Replace the cap on the lancing
device. Make sure the notch on the
cap lines up with the notch on the
lancing device.
6
7
Adjust the lancet depth to a
comfortable penetration level. Start
at 2. For tougher skin, dial to a
higher number.
Press the priming button in as
far as it will go, like a pen. Do not
press the release button while
pressing the priming button. The
release button turns yellow when
the lancing device is ready.
8
Set the lancing device aside until
you are ready to obtain a blood
sample.
Wash and dry your hands before
you perform a blood glucose test.
1
Check the use by date on the test
strip container. Do not use test
strips past the use by date.
2
Remove a test strip from the test
strip container. Close the cap
tightly.
3
Insert the test strip into the meter
in the direction of the arrows until
the meter beeps.
4
A flashing blood drop symbol
appears.
Setting up the Lancing Device
Performing a Blood Glucose Test
9
Hold the lancing device firmly
against the edge of your fingertip
and press the yellow release
button to poke your fingertip.
Touch the end of the test strip to
the blood drop. Do not put blood on
top of the test strip.
The meter beeps and flashes
when there is enough blood in the
test strip.
The test result appears on the
display after 5 seconds.
Remove the lancing device cap.
Point the lancet away from you.
Slide out the ejector to release the
lancet.
Ejector
Release button
Cap
Priming
button
Error Messages
5
6
NOTE
Contact Accu-Chek Customer Care at 1-800-363-7949 if the
error message continues.
ACCU-CHEK, ACCU-CHEK AVIVA, and SOFTCLIX are
trademarks of Roche.
© 2014 Roche Diagnostics
07022425001-0114
LAST UPDATE: 2013-04
Twist off the lancet’s protective
cover.
Before you finish…
Penetration depth
1
2
3
4
5
6
1
WARNING
WARNING
This Quick Start Guide does not replace the User’s
Manual for your Accu-Chek Aviva blood glucose
meter. The User’s Manual contains important
handling instructions and additional information.
The lancing device is for single patient use only and
should not be shared.