Optics Alignment (Required for Periodic Inspection or Installing X-Ray Tube)
MiniFlex 600: Benchtop X-ray Diffractometer
37
9
Select Meas. for Axis control, enter the conditions shown in the figure
below, and click the Scan button to perform measurement of the direct
beam peak position (Full Width at Half Maximum: FWHM).
Scan conditions, Absolute, Start: -0.4, Stop: 0.4, Step: 0.01, Speed: 5, Scan
mode: Continuous
Count mode under Detector: Int.
Fig. 6.2.11 Theta/2-theta axis scan conditions
(rough adjustment stage)
Fig. 6.2.12 Direct beam measurement result (rough adjustment)
By clicking the Peak position calculation button, the peak position can be
calculated. Specify
Full Width at Half Maximum
for
Peak Search
Method
.
Peak position
calculation
Peak angle,
intensity and
FWHM