Installing the DS and IHS (Length limiting)
MiniFlex 600: Benchtop X-ray Diffractometer
27
Installing the DS and IHS (Length limiting)
1
Install the DS=1.25° to the DS section.
2
Insert the IHS layered with DS of X-ray tube side.
The IHS is a slit to limit the longitude direction of a slit.
During a variable slit mode, insert only the IHS.
Fig. 5.23 Enlarged image of DS
Connecting the power cable
1
Connect the power cable to the power cable connector on the rear side
of MiniFlex 600 main unit.
For power supply input cable with plug (single phase 100 V – 240 V AC
power supply), connect the power supply input cable to a dedicated power
supply of 15 A grounded power outlet (3-contact). The standard included
power cable is compatible up to 125 V AC.
Connecting with Computer
1
Connect the LAN connection port of the computer to the LAN
connection port on the rear side of MiniFlex 600 main unit (inner side)
with the LAN cable.
If there are more than two LAN connection ports on the PC side, connect to
the specified LAN connection port.
Position to insert DS and IHS