Setting and Adjustment
34
MiniFlex 600: Benchtop X-ray Diffractometer
Normally, the minimum angle is set to 2
°
to prevent the direct beam from
entering the detector accidentally.
4
Confirm the angle correction and slit correction.
①
Select
Options
-
Angle Correction
. Set the angle correction to
Disable.
②
Select
Options
-
Slit Correction
. Set the slit correction to
Disable.
Fig. 6.2.7 Angle Correction window
Fig. 6.2.8 Slit correction window
When
Manual Control
is already opened, this change will not be updated to
system settings. Close the
Manual Control
window, and then re-open it.
5
When aligning the optics for the first time after installing the system,
select Options - Detector Setup – SC-70, and enter 500 mV in the PHA
baseline box, 1000 mV in the PHA window box, and the HV value
indicated on the label pasted on SC (see Fig. 5.10) in the HV box as the
initial value.