Setting and Adjustment
60
MiniFlex 600: Benchtop X-ray Diffractometer
⑥
Click the
Execute
button.
The correction value is registered in the MiniFlex Guidance
automatically after the measurement is completed.
⑦
Click the
OK
button.
Fig. 6.2.41 Measurement condition
14
Confirm that the direct beam is blocked.
①
Mount the Cu absorber (0.3 mm x 1) on the Soller slit box. If the
beam stop is not mounted, mount it.
②
E
xecute measurement of direct beam and confirm that the direct beam
is blocked by the beam stop.
Scan conditions, Absolute, Start: -0.4, Stop: 0.4, Step: 0.01, Speed: 5, Scan
mode: Continuous
Count mode under Detector: Int.
・
If part of or the whole direct beam can be seen, perform the procedures
in
15
Adjust the beam stop
.
・
When it is confirmed that the direct beam is blocked by the beam stop,
go on to
16
Perform the beam stop position confirmation
measurement.
Enter 0° when using the
direct beam