Installing the Scintillation Counter (SC)
MiniFlex 600: Benchtop X-ray Diffractometer
19
CAUTION
Mistaking the IN and OUT connection of the cooling water degrades the
cooling efficiency and may damage the X-ray tube.
Installing the Scintillation Counter (SC)
1
Remove the parts for transportation (see Fig. 5.4a and Fig. 5.4b).
Fig. 5.4a Removing the parts for transportation (for SC)
Fig. 5.4b Removing the parts for transportation (with beam stop)
・
Before installing the SC, remove the parts for transportation, rotate the
balancer of driving motor by hand, and move the goniometer arm to a
position raised approx. 90° from the horizontal position. The SC can be
installed by inserting a hex wrench from the hole on the cover at the rear side
of the goniometer arm.
・
There are two types of parts for transportation. The type differs depending on
the system configuration (see Fig. 5.4a and Fig. 5.4b).