Optics Alignment (Required for Periodic Inspection or Installing X-Ray Tube)
MiniFlex 600: Benchtop X-ray Diffractometer
35
Fig. 6.2.9 Detector Setup window
6
Select Control – Initialize Individual Axis – Theta/2-Theta, and click
the Execute button. In the same manner, select Control - Initialize
Individual Axis – Detector # 1, and click the Execute button.
7
Select Control - XG Control, and click the ON button of X-Ray. Set the
tube voltage and tube current to 30 kV - 10 mA respectively.
Fig. 6.2.10 XG Control window
The environment for optics alignment has been set up by the above
procedures.
8
Adjust the X-ray tube in vertical direction to position the direct beam
at an angle of 2θ/θ= 0°.
①
Select
Control
-
Manual Control
.