Installation of MiniFlex600
18
MiniFlex 600: Benchtop X-ray Diffractometer
③
Confirm that the O-rings are placed. Install the X-ray tube into the
tube shield, making sure that the tube positioning pin fits the tube
shield pin hole (see Fig. 5.2 (b)). Tighten the two tube fixing screws
and fix the X-ray tube into position.
(a) Tube shield (without the X-ray tube)
(b) After the X-ray tube is installed
Fig. 5.2 Installing the X-ray tube
Connecting the Cooling Water Hose
1
Connect the cooling water hoses included with the equipment to the IN
and OUT ports (see Fig 5.3) on the rear side of the main unit.
Couple the connection ports securely using the included AMC couplings to
prevent water leakage.
2
Connect the main unit and the cooling water supply source (either tap
water or heat exchanger) using the cooling water hoses included with
the equipment, which should be cut to an adequate length.
Couple the connection ports securely using the cooling-water hose bands
(for tap water) or AMC couplings (for heat exchanger) to prevent water
leakage.
3
Attach the IN and OUT labels on the end of the cooling hoses.
Fig 5.3 Connecting the Cooling Water Hoses
O-rings
Tube positioning
pin
Tube fixing
screws
MiniFlex 600 (rear side
of main body)
OUT IN
Cooling water
supply source
Cooling water hoses