Setting and Adjustment
56
MiniFlex 600: Benchtop X-ray Diffractometer
7
Select Meas. for Axis control, enter the conditions shown in the figure
below, and click the Scan button to perform measurement of the direct
beam peak position (Full Width at Half Maximum: FWHM).
Scan conditions, Absolute, Start: -0.4, Stop: 0.4, Step: 0.01, Speed: 5, Scan
mode: Continuous
Count mode under Detector: Int.
Fig. 6.2.36 Theta/2-theta axis scan conditions
(rough adjustment stage)
Fig. 6.2.37 Direct beam measurement result (rough adjustment)
By clicking the Peak position calculation button, the peak position can be
calculated. Specify the Full Width at Half Maximum for the Peak Search
Method.
When the precision measurement result shows a peak angle within the
range of ± 0.01°, the adjustment is completed.
Peak position
calculation
Peak angle and
intensity and
FWHM