Setting and Adjustment
38
MiniFlex 600: Benchtop X-ray Diffractometer
When the precision measurement result shows a peak angle within the
range of ± 0.01°, the adjustment is completed.
If it is 0.01° or more, adjust the position of the X-ray tube.
If the peak cannot be determined, broaden the range of start and stop of
scan condition and try it again.
10
Adjust the vertical position of the X-ray tube.
①
When the sample chamber door is opened, a thick adjustment screw,
with thin locking screws on its left and right side can be seen in the
center of the rear side of the tube shield.
Fig. 6.2.13 Tube position vertical adjustment mechanism
②
If the peak position is off to the positive side, loosen the locking
screws and turn the tube position vertical adjustment screw in
clockwise direction (slightly).
③
If the peak position is off to the negative side, turn the tube position
vertical adjustment screw in counterclockwise direction (slightly).
After the adjustment is completed, tighten the locking screws and
perform the measurement of direct beam peak position again by
procedures in
8
.
Locking screws (two)
(Lock by rotating in clockwise
direction)
Tube position vertical adjustment screw
Rotate in clockwise direction: Peak moves
to minus direction
Rotate in counter-clockwise direction: Peak
moves to plus directon
*
Rotation of 10°
=
0.01°