Startup and Shutdown (Daily routine)
78
MiniFlex 600: Benchtop X-ray Diffractometer
If the HV ENABLE key is not inserted, X-ray cannot be generated.
The startup of MiniFlex 600 is now completed by the procedures above.
Shutting Down the System
1
Confirm that the X-ray generation has stopped.
Click Control – XG Control and then click the OFF button of X-Ray in
MiniFlex Guidance to stop the X-ray generation.
When the X-ray generation stops, the yellow X-ray warning lamp on top of
the enclosure will turn off.
2
End MiniFlex Guidance. (Each window can be left open)
3
Shut down the PC.
4
Wait at least 3 or more minutes after the X-ray generation has stopped,
and then press the Power OFF switch (see Fig. 7.1).
5
Turn OFF the “Power circuit breaker” (see Fig. 2.2) on the rear side of
MiniFlex 600 main unit.
6
When a heat exchanger is used, turn OFF the power of the heat
exchanger.
When using tap water, close the faucet of tap water and stop the water from
running.
If you store the HV ENABLE key in a separate place, remove the
key from the panel.
The shutting down of MiniFlex 600 is now completed by the procedures above.