Installation of MiniFlex600
Unpacking
Unpack the main unit and the component parts, and verify that they match with
the list of shipped items (see “
”).
Installing the X-Ray tube
1
Remove the upper cover of the enclosure.
①
Remove the six fixing screws on (3 x 2) on both sides of the main
unit.
②
Pull the cover slightly backward, and then lift it up.
Fig. 5.1 Removing the upper cover of the enclosure
2
Install the X-ray tube.
①
Remove the upper cover. The X-ray tube is on the left side when
looking at the equipment from the front side.
②
Place the O-rings (two, included within the storage box) on the upper
side of the tube shield.
Without the O-rings, the cooling water may leak out.
Upper side of the
enclosure
Direction to remove
cover
Screws
fastening the
upper cover
(Total of six
screws on left