Setting and Adjustment
70
MiniFlex 600: Benchtop X-ray Diffractometer
Fluorescent X-ray Reduce High = Standard High
PHA baseline = Fluorescent X-ray Reduce Low
PHA window = Fluorescent X-ray Reduce High - Fluorescent
X-ray Reduce Low
Pseudo SC:
A setting to take in 2 extra channels (high energy side and
low energy side) compared with standard mode. Intensity increases
compared with Standard, but becomes more easily affected by
fluorescent X-rays.
How to calculate from Standard is the following:
Pseudo SC Low = Standard Low +2
Pseudo SC High = Standard High +2
PHA baseline = Pseudo SC Low
PHA window = Pseudo SC High - Pseudo SC Low