Optics Alignment (Required for Periodic Inspection or Installing X-Ray Tube)
MiniFlex 600: Benchtop X-ray Diffractometer
63
Fig. 6.2.46 Measurement result of beam stop position confirmation
(adjustment completed)
17
Tighten the beam stop fixing screw and beam stop mounting screw
The adjustment of D/teX Ultra is completed by the above procedures.
2°