Maintenance
92
MiniFlex 600: Benchtop X-ray Diffractometer
remaining in the hose can be prevented from leaking out by keeping the end
of the removed cooling water hose at a high position.
4
Remove the upper cover
①
of the enclosure. (Tool used: Philips
screwdriver)
Fig. 9.1 Separating upper cover and separating X-ray tube
5
Loosen the two screws
②
of X-ray tube. (Tool used: Slotted
screwdriver)
6
Remove the X-ray tube
③
.
CAUTION
Make sure that the water remaining inside the X-ray tube does not drip inside
the tube shield. If the water drips into the X-ray tube shield, thoroughly dry the
interior by absorbing the water with a piece of cloth or soft paper.
Fig. 9.2 Inside the enclosure (rear view with the upper cover removed)
Sample
chamber
door
X-ray tube
Sample
position
Tube shield
Tube fixing screws
Standard sample
holder
Counter (SC)
Cooling water tube
connection port
①