Installing the Scintillation Counter (SC)
MiniFlex 600: Benchtop X-ray Diffractometer
21
Install the receiving slit box so it is positioned the nearest to the SC side with
the SC removed. If the position is incorrect, measurements cannot be performed
correctly.
4
Connect the SC cable to the cable joint connector on the SC. (see Fig.
5.8)
Fig 5.8 SC
Fig 5.9 Assembling the SC
A label is attached on the rear side of the SC. The HV value indicated on the
label will be used as the initial value upon the adjustment of HV value. Before
mounting the upper cover of the enclosure, take note and remember this value.
Fig 5.10 Label attached on SC
Cable joint
connector
Window
SC cable
SC fixing screws
This HV value will be used
as the initial HV value of the
SC.