Optics Alignment (Required for Periodic Inspection or Installing X-Ray Tube)
MiniFlex 600: Benchtop X-ray Diffractometer
61
Fig. 6.2.42 Conditions of direct beam confirmation measurement
15
Adjust the beam stop
①
Adjust the beam stop with the adjustment screw.
Loosen the beam stop mounting screw and beam stop fixing screw.
Then, turn the adjustment screw to adjust the height of the beam stop.
②
Perform the procedures in
14
Confirm that the direct beam is
blocked
again and confirm that the direct beam is blocked by the
beam stop.
Fig. 6.2.43 Adjusting the beam stop
16
Perform the beam stop position confirmation measurement.
CAUTION
Before performing the beam stop position confirmation measurement, make
sure that the direct beam is blocked by the beam stop in
14
Confirm that the
direct beam is blocked.
The detector may be damaged if the direct beam enters
the D/teX Ultra directly.
Beam stop
fixing screw
Beam stop
mounting
screw
Adjusting
screw