Installing the D/teX Ultra (Optional)
MiniFlex 600: Benchtop X-ray Diffractometer
23
Fig. 5.13 Moving the receiving slit box
5
Connect the power supply cable and communication LAN cable to
each connector of the detector.
Fix the SC cable so it will not collide with the goniometer arm.
Fig. 5.14 Installing the D/teX Ultra Fig. 5.15 SC cable
A label is attached on D/teX Ultra. The discriminator value indicated on the
label is used as the Low and High discriminator value. Take note and remember
these values before mounting the upper cover of the enclosure.
For the figure below, the discriminator value is indicated as Low=37 and
High=53. Record as BL=Low=37, WD=High(53)-Low(37)=16.
Fig. 5.16 Label attached on D/teX Ultra
Receiving slit box
fixing screws
Slide to the
D/teX Ultra side
Power
supply cable
LAN cable
SC Cable