Installation of MiniFlex600
22
MiniFlex 600: Benchtop X-ray Diffractometer
Installing the D/teX Ultra (Optional)
1
If the SC is installed, remove it.
2
To install the D/teX Ultra, move the position of the receiving slit box to
the sample side.
3
Install the D/teX Ultra to the goniometer arm with the two receiving
slit box fixing screws. (Fig. 5.13)
4
Move the receiving slit box so it is positioned right next to the D/teX
Ultra (Fig. 5.13).
Rotate the balancer of driving motor by hand, and move the goniometer arm to
a position raised approx. 90° from the horizontal position. Remove the SC by
inserting a hex wrench from the hole on the cover at the rear side of the
goniometer arm, and install the D/teX Ultra at this position.
Fig. 5.11 Position to install D/teX Ultra
Fig. 5.12 Installing the D/tex Ultra