Setting and Adjustment
44
MiniFlex 600: Benchtop X-ray Diffractometer
Fig. 6.2.20 Detector Setup window
4
Select Control - Initialize Individual Axis - Theta/2-Theta and click the
Execute button. In the same manner, select Control - Initialize
Individual Axis – Detector # 1, and click the Execute button.
Next, execute the D/teX Detector Center Correction. To execute with direct
beam, move on to
5
. To execute with a sample, move on to
6
. It is
recommended to execute by the direct beam.
5
Execute the D/teX Detector Center Correction. - with direct beam
①
Select
Tasks
-
Package Measurement
, -
Maintenance
-
D/teX
Detector Adjustment - D/teX Detector Center Correction
.
②
Check the
Center Correction Clear
checkbox.
③
Click the
Execute
button.
④
Uncheck the
Center Correction Clear
checkbox.
⑤
Check the
Theta/2-theta scan conditions
checkbox.
⑥
Set DS=1.25°, absorber (Cu plate 0.3 mm, 1 or 2 plates), and RS =
OPEN. Enter 0° in the
Theta/2-theta position (deg)
box.
⑦
Click the
Execute
button.
The correction value is registered in the MiniFlex Guidance
automatically after the measurement is completed.
⑧
Click the
OK
button.
The correction value can be seen on the History window. (Click the History
command under the View menu to display the History window.)