Optics Alignment (Required for Periodic Inspection or Installing X-Ray Tube)
MiniFlex 600: Benchtop X-ray Diffractometer
39
11
When the peak position comes near the range of ±0.01°, change the
measurement conditions to those in Fig. 6.2.14 and perform precision
measurement of the peak position.
Scan conditions, Absolute, Start: -0.2, Stop: 0.2, Step: 0.01, Speed: 2, Scan
mode: Continuous
Count mode under Detector: Int.
When the direct beam position is within the range of ±0.01°,the tube
position adjustment is completed.
Fig. 6.2.14 Direct beam measurement conditions
(precision measurement)
Fig. 6.2.15 Direct beam measurement result (precision measurement)
within ±0.01°