Setting and Adjustment
58
MiniFlex 600: Benchtop X-ray Diffractometer
Fig. 6.2.39 Direct beam measurement conditions
(precision measurement)
Fig. 6.2.40 Direct beam measurement result (precision measurement)
10
In general, correction value is not registered in MiniFlex 600. The
angle error can be minimized by setting the “
”, which will be described later.
・
MiniFlex 600 performs measurement by reflection method (measurement of
diffraction beam from the sample surface) in general. With this system, about
one-third of the direct beam angle error is the angle error (shift) associated
with the reflection method measurement. The angle error is about 0.003° for
0.01°, and correction is unnecessary.
・
Initialization procedure: Function to move the goniometer to DATUM angle
within ±0.01°