Setting and Adjustment
62
MiniFlex 600: Benchtop X-ray Diffractometer
①
Remove the Cu absorber mounted on Soller slit box.
②
Set the setting jig (with Si) on the sample holder.
③
Check that the beam stop is not set at an extreme position.
④
Confirm that the background intensity at 2θ=2° or larger is not
blocked by the beam stop.
⑤
If an adjustment is required again, return to
15
Adjust the beam
stop
and perform the adjustment.
Scan conditions, Absolute, Start: 1, Stop: 5, Step: 0.01, Speed: 10, Scan
mode: Continuous
Count mode under Detector: Diff.
Fig. 6.2.44 Conditions of beam stop position confirmation
measurement
Fig. 6.2.45
Measurement result of beam stop position confirmation
(insufficient adjustment)
3°