MiniFlex 600: Benchtop X-ray Diffractometer
13
Installation Condition
To install MiniFlex 600, the following conditions must be satisfied:
Installation Room
The MiniFlex 600 is a benchtop installation type equipment. Prepare a bench or
a desk with the space and load limit specified below, and install it on a stable
bench.
Dimensions and Weight of the Equipment
560 (W) x 460 (D) x 700 mm (H), weight: approx. 80 kg
・
Additional space for a computer and printer is required.
・
Do not place any object within 30 cm from the vent located on the right side
of MiniFlex 600.
Environmental Conditions of Installation Room
Vibration of floor:
The floor must not have vibrations perceivable to
the human body.
Temperature:
15°C to 25°C
Relative humidity:
Below 65% (no condensation)
Abrupt changes in temperature and relative
humidity adversely affects data measurement
accuracy, as well as the performance and service
life of the equipment.
Magnetic field:
There must be no source of magnetic fields nearby
that may adversely affect the LCD display of the
computer.
Others:
・
The room must be well-ventilated, without
direct sunlight or heat radiation from
high-temperature objects.
・
The atmosphere must not contain inflammable
gases, corrosive gases, or salinity.
・
The room must not be subject to the infiltration
of cotton dust, sand, soil dust, metallic dust, or
other dust.