MiniFlex 600: Benchtop X-ray Diffractometer
9
Specifications
X-Ray Generator
Tube output voltage:
20 to 40 kV (1 kV step)
Tube output current:
2 to 15 mA (1 mA step)
High-voltage generation method:
High-frequency Cockcroft-Walton method
Stability:
Within ±0.05% for both the tube voltage and tube
current, with reference to ±10% of input power
variation
Tube shield
Applicable X-ray tube: Toshiba A-21, A-41, PANalytical PW22XX, etc.
Focusing-position adjustment mechanism:
±1 mm, continuously variable through the use of a
spring plate
X-ray shutter:
Mechanical rotary shutter
K
β
filter:
Ni-filter (for Cu-tube, standard thickness 0.015
mm). Other filters are optional.
Goniometer (Angle measuring device)
Type:
Vertical θ/2θ
Goniometer radius:
150 mm
Scanning axis:
θ/2θ linked
2θ scanning range:
−
3° to +145°
2θ measurement range: +2° to +145°
θ/2θ axis minimum step angle:
0.005° (2θ)
Positioning speed:
500° / min (2θ)
Scanning speed:
0.01 to 100° / min.
Datum angle:
2θ = 10°
Slit:
DS
1.25°, 0.625°
IHS (length limiting) 10 mm, 5 mm (optional), 2 mm (optional)
Auxiliary DS
This is a variable slit to limit the width of X-ray
beam irradiated on sample to be approx. 20 mm
(sample width) at low angle area
SS
1.25°, 0.625° (optional), 4.2° (optional)