Setting and Adjustment
54
MiniFlex 600: Benchtop X-ray Diffractometer
Fig. 6.2.34 Detector Setup window
4
Select Control - Initialize Individual Axis - Theta/2-Theta and click the
Execute button. In the same manner, select Control - Initialize
Individual Axis - Detector # 1, and click the Execute button
When
Manual Control
is already opened, this change will not be updated to
system settings. Close the
Manual Control
window, and then re-open it.
5
Clear the D/teX Detector Center Correction.
①
Select
Tasks
-
Package Measurement
-
Maintenance -
D/teX
Detector Adjustment - D/teX Detector Center Correction
.
②
Check the
Center Correction Clear
checkbox.
③
Click the
Execute
button.
④
Close the window.
The environment for optics alignment has been set up by the above procedures.