Description of Device Parameters

Proline Cubemass 300

Modbus RS485

Coriolis flowmeter

Menu/Variable

Proline Device

Access status tooling:

Operation

Setup

Diagnostics

Locking status:

Access status tooling:

Enter access code:

System

Sensor

Input

Output

Communication

Application

Diagnostics

P

P

P

P

Expert

GP01088D/06/EN/02.17

71376168
Valid as of version

01.01.zz (Device firmware)

Содержание Proline Cubemass 300

Страница 1: ...is flowmeter Menu Variable Proline Device Access status tooling Operation Setup Diagnostics Locking status Access status tooling Enter access code System Sensor Input Output Communication Application...

Страница 2: ......

Страница 3: ...2 Modbus info submenu 165 3 6 3 Modbus data map submenu 166 3 6 4 Web server submenu 167 3 6 5 WLAN settings submenu 170 3 7 Application submenu 176 3 7 1 Totalizer 1 to n submenu 177 3 7 2 Concentra...

Страница 4: ...ice Commissioning measurements under difficult conditions Optimal adaptation of the measurement to difficult conditions Detailed configuration of the communication interface Error diagnostics in diffi...

Страница 5: ...Operating menu for operators and maintenances Parameter n Operator Maintenance Task oriented Function orien ted Expert Wizard 1 Parameter 1 Wizard n Parameter n Parameter n Input A0029160 EN 1 Sample...

Страница 6: ...arameter is only available under these specific conditions Description Description of the parameter function Selection List of the individual options for the parameter Option 1 Option 2 User entry Inp...

Страница 7: ...structions Measuring device Documentation code Cubemass C 300 BA01494D 1 5 2 Supplementary device dependent documentation Special documentation Contents Documentation code Information on the Pressure...

Страница 8: ...ates where the associated description of the submenu or parameter can be found Expert Direct access 0106 11 Locking status 0004 12 Access status 0005 13 Ent access code 0003 13 System 13 Display 14 Co...

Страница 9: ...tatus input 1 to n 106 Output 108 Curr output 1 to n 109 PFS output 1 to n 124 Relay output 1 to n 148 Double pulse out 155 Communication 160 Modbus config 160 Modbus info 165 Modbus data map 166 Web...

Страница 10: ...5 10 Endress Hauser Operating time 0652 185 Diagnostic list 186 Event logbook 190 Device info 192 Mainboard module 196 Sens electronic 197 I O module 1 198 I O module 2 198 I O module 3 200 Display mo...

Страница 11: ...sor 46 I O config 101 Input 103 Output 108 Communication 160 Application 176 Diagnostics 183 Direct access Navigation Expert Direct access 0106 Description Use this function to enter the access code t...

Страница 12: ...wo or more types of write protection are active the write protection with the highest priority is shown on the local display In the operating tool all active types of write protection are displayed De...

Страница 13: ...parameter 13 If additional write protection is active this restricts the current access authorization even further Display Detailed information on access authorization is provided in the User roles a...

Страница 14: ...0107 18 0 bargraph 1 0123 19 100 bargraph 1 0125 20 Decimal places 1 0095 20 Value 2 display 0108 20 Decimal places 2 0117 22 Value 3 display 0110 22 0 bargraph 3 0124 23 100 bargraph 3 0126 23 Decima...

Страница 15: ...s Portuguesa Polski Ru Svenska T rk e Chinese Japanese Korean Bahasa Indonesia ti ng Vi t Vit e tina Czech Factory setting English alternatively the ordered language is preset in the device Format dis...

Страница 16: ...onfigured This setting only applies to normal operation The Value 1 display parameter 18 to Value 4 display parameter 24 are used to specify which measured values are shown on the local display and in...

Страница 17: ...1 value max option XXXX XXX X X kg h 900 00 A0013099 Bargr 1 value option XXXX XXX X X 900 00 900 00 kg h l h A0013098 2 values option XXXX XXX X X kg h 900 00 l h 900 00 A0013100 Val large 2val optio...

Страница 18: ...Carrier vol fl Targ corr vol fl Carr corr vol fl Density Ref density Ref dens altern GSV flow GSVa NSV flow NSVa S W volume flow Water cut Oil density Water density Oil mass flow Water mass flow Oil v...

Страница 19: ...100 under optimum conditions Oscil damping option Displays the current oscillation damping Oscillation damping is an indicator of the sensor s current need for excitation power Signal asymmetry optio...

Страница 20: ...try The unit of the displayed measured value is taken from the System units submenu 59 Decimal places 1 Navigation Expert System Display Decimal places 1 0095 Prerequisite A measured value is specifie...

Страница 21: ...nsity Ref density Ref dens altern GSV flow GSVa NSV flow NSVa S W volume flow Water cut Oil density Water density Oil mass flow Water mass flow Oil volume flow Water corr v fl Oil corr vol fl Water vo...

Страница 22: ...uisite A measured value is specified in the Value 2 display parameter 20 Description Use this function to select the number of decimal places for measured value 2 Selection x x x x xx x xxx x xxxx Fac...

Страница 23: ...made in the Value 3 display parameter 22 Description Use this function to enter the 0 bar graph value to be shown on the display for the measured value 3 User entry Signed floating point number Facto...

Страница 24: ...this function to select the number of decimal places for measured value 3 Selection x x x x xx x xxx x xxxx Factory setting x xx Additional information Description This setting does not affect the mea...

Страница 25: ...rerequisite A measured value is specified in the Value 4 display parameter 24 Description Use this function to select the number of decimal places for measured value 4 Selection x x x x xx x xxx x xxx...

Страница 26: ...ant for the reaction time of the local display to fluctuations in the measured value caused by process conditions User entry 0 0 to 999 9 s Factory setting 0 0 s Additional information User entry Use...

Страница 27: ...er 26 the Free text option is selected Description Use this function to enter a customer specific text for the header of the local display User entry Max 12 characters such as letters numbers or speci...

Страница 28: ...ntry 20 to 80 Factory setting Depends on the display Backlight Navigation Expert System Display Backlight 0111 Prerequisite One of the following conditions is met Order code for Display operation opti...

Страница 29: ...Additional information User interface The maximum number of days is 9999 which is equivalent to 27 years Last backup Navigation Expert System Config backup Last backup 2757 Description Displays the t...

Страница 30: ...ration saved in the device memory is compared with the current device configuration of the HistoROM backup The following message appears on local display Comparing files The result can be viewed in Co...

Страница 31: ...ckup copy in the device memory The settings for the transmitter are not identical Set not ident The current device configuration of the HistoROM is not identical to the backup copy in the device memor...

Страница 32: ...ostic behavior at the factory The user can change this assignment for specific diagnostic information in the Diagn behavior submenu 32 The following options are available in the Assign behavior of dia...

Страница 33: ...0710 35 Diagnostic no 302 0739 35 Diagnostic no 441 0657 35 Diagnostic no 442 0658 36 Diagnostic no 443 0659 36 Diagnostic no 444 0740 36 Diagnostic no 543 0643 37 Diagnostic no 830 0800 37 Diagnosti...

Страница 34: ...Expert System Diagn handling Diagn behavior Diagnostic no 140 0708 Description Use this function to change the diagnostic behavior of the diagnostic message 140 Sensor sig asym Selection Off Alarm War...

Страница 35: ...tion Expert System Diagn handling Diagn behavior Diagnostic no 302 0739 Description Use this function to change the diagnostic behavior of the diagnostic message 302 Verific active Selection Alarm War...

Страница 36: ...32 Diagnostic no 443 Pulse output 1 to n Navigation Expert System Diagn handling Diagn behavior Diagnostic no 443 0659 Prerequisite The measuring device has a pulse frequency switch output Description...

Страница 37: ...ge the diagnostic behavior of the diagnostic message 543 Double pulse out Selection Off Alarm Warning Logbook only Factory setting Warning Additional information For a detailed description of the opti...

Страница 38: ...System Diagn handling Diagn behavior Diagnostic no 832 0681 Description Use this function to change the diagnostic behavior of the diagnostic message 832 Electronic temp Selection Off Alarm Warning L...

Страница 39: ...tion Expert System Diagn handling Diagn behavior Diagnostic no 835 0702 Description Use this function to change the diagnostic behavior of the diagnostic message 835 Process temp Selection Off Alarm W...

Страница 40: ...tion Expert System Diagn handling Diagn behavior Diagnostic no 913 0712 Description Option for changing the diagnostic behavior of the diagnostic message 913 Medium unsuitab Selection Off Alarm Warnin...

Страница 41: ...ation For a detailed description of the options available see 32 3 1 4 Administration submenu Navigation Expert System Administration Administration Def access code 41 Reset acc code 43 Device reset 0...

Страница 42: ...local display the symbol in front of a parameter indicates that the parameter is write protected The parameters that cannot be write accessed are grayed out in the Web browser Once the access code ha...

Страница 43: ...tes m and seconds s Additional information User interface The maximum number of days is 9999 which is equivalent to 27 years Reset acc code Navigation Expert System Administration Reset acc code Reset...

Страница 44: ...the factory setting Restart device The restart resets every parameter whose data are in the volatile memory RAM to the factory setting e g measured value data The device configuration remains unchange...

Страница 45: ...If the new software option is not displayed or all software options have been deleted the code entered was either incorrect or invalid If the code entered is incorrect or invalid enter the old activat...

Страница 46: ...option and HBT Monitoring option Order code for Application package option EB Heartbeat Verification Monitoring Concentration option Order code for Application package option ED Concentration 3 2 Sens...

Страница 47: ...ect vol flow 1851 48 Density 1850 48 Ref density 1852 49 Temperature 1853 49 Pressure value 6129 49 Concentration 1887 49 Target mass flow 1864 50 Carrier mass fl 1865 50 Targ corr vol fl 1893 51 Carr...

Страница 48: ...red and the density currently measured Dependency The unit is taken from the Volume flow unit parameter 61 Correct vol flow Navigation Expert Sensor Measured val Process variab Correct vol flow 1851 D...

Страница 49: ...rrently measured User interface Signed floating point number Additional information Dependency The unit is taken from the Temperature unit parameter 67 Pressure value Navigation Expert Sensor Measured...

Страница 50: ...e SW option overv parameter 45 Description Displays the mass flow currently measured for the target medium User interface Signed floating point number Additional information Dependency The unit is tak...

Страница 51: ...rr vol fl 1894 Prerequisite With the following conditions Order code for Application package option ED Concentration The WT option or the User conc option is selected in the Concentr unit parameter Th...

Страница 52: ...bled are displayed in the SW option overv parameter 45 Description Displays the volume flow currently measured for the carrier fluid User interface Signed floating point number Additional information...

Страница 53: ...w 1 parameter 1 107 1 overflow 10000000 m3 Current totalizer reading 11968457 m3 Tot overflow 1 to n Navigation Expert Sensor Measured val Totalizer Tot overflow 1 to n 0910 1 to n Prerequisite A proc...

Страница 54: ...n Current input 1 to n Measured val 1 to n 1603 1 to n 54 Measur curr 1 to n 1604 1 to n 54 Measured val 1 to n Navigation Expert Sensor Measured val Input values Current input 1 to n Measured val 1...

Страница 55: ...s Val stat inp 1 to n Val stat inp 1353 1 to n Description Displays the current input signal level User interface High Low Output values submenu Navigation Expert Sensor Measured val Output values Out...

Страница 56: ...his function to display the actual measured value of the output current User interface 0 to 30 mA Pulse frequency switch output 1 to n submenu Navigation Expert Sensor Measured val Output values PFS o...

Страница 57: ...onductive for the duration of the pulse NO contact and is safety oriented NC NO t t 0 1 0 1 A0028726 0 Non conductive 1 Conductive NC NC contact normally closed NO NO contact normally open The output...

Страница 58: ...cles 0815 1 to n 58 Max cycles no 0817 1 to n 59 Switch status Navigation Expert Sensor Measured val Output values Relay output 1 to n Switch status 0801 1 to n Description Displays the current status...

Страница 59: ...asured val Output values Double pulse out Double pulse out Pulse output 0987 59 Pulse output Navigation Expert Sensor Measured val Output values Double pulse out Pulse output 0987 Description Displays...

Страница 60: ...unit 0557 67 Pressure unit 0564 67 Date time format 2812 68 Mass flow unit Navigation Expert Sensor System units Mass flow unit 0554 Description Use this function to select the unit for the mass flow...

Страница 61: ...Mass text parameter 70 Mass unit Navigation Expert Sensor System units Mass unit 0574 Description Use this function to select the unit for the mass Selection SI units g kg t US units oz lb STon Facto...

Страница 62: ...kgal min us kgal h us kgal d us Mgal s us Mgal min us Mgal h us Mgal d us bbl s us oil bbl min us oil bbl h us oil bbl d us oil bbl s us tank bbl min us tank bbl h us tank bbl d us tank Imperial unit...

Страница 63: ...Expert Sensor System units Volume unit 0563 Description Use this function to select the unit for the volume Selection SI units cm dm m ml l hl Ml Mega US units af ft fl oz us gal us kgal us Mgal us bb...

Страница 64: ...Sbbl h us liq Sbbl d us liq Imperial units Sgal s imp Sgal min imp Sgal h imp Sgal d imp Factory setting Country specific Nl h Sft min Additional information Result The selected unit applies for Corre...

Страница 65: ...o select the unit for the density Selection SI units g cm g m g ml kg l kg dm kg m SD4 C SD15 C SD20 C SG4 C SG15 C SG20 C US units lb ft lb gal us lb bbl us oil lb bbl us tank Imperial units lb gal i...

Страница 66: ...rature of 4 C 39 F 15 C 59 F 20 C 68 F For an explanation of the abbreviated units 232 Customer specific units The unit for the customer specific density is specified in the Density text parameter 73...

Страница 67: ...value parameter 212 Minimum value parameter 212 Maximum value parameter 213 Minimum value parameter 213 Maximum value parameter 214 Minimum value parameter 214 External temp parameter 89 Temperature...

Страница 68: ...ts Date time format 2812 Description Use this function to select the desired time format for calibration history Selection dd mm yy hh mm dd mm yy am pm mm dd yy hh mm mm dd yy am pm Factory setting d...

Страница 69: ...74 Energy offset 0599 75 Energy factor 0586 75 Pressure text 0581 75 Pressure offset 0580 76 Pressure factor 0579 76 Volume text Navigation Expert Sensor System units User spec units Volume text 0567...

Страница 70: ...apting the user specific volume unit and volume flow unit without time User entry Signed floating point number Factory setting 0 Additional information Description Value in user specific unit factor v...

Страница 71: ...in CENT h CENT d Mass offset Navigation Expert Sensor System units User spec units Mass offset 0562 Description Use this function to enter the zero point shift for the user specific mass and mass flow...

Страница 72: ...orr vol unit parameter 64 Example If the text GLAS is entered the choose list of the Cor volflow unit parameter 64 shows the following options GLAS s GLAS min GLAS h GLAS d Corr vol offset Navigation...

Страница 73: ...liter Density offset Navigation Expert Sensor System units User spec units Density offset 0571 Description Use this function to enter the zero point shift for the user specific density unit Value in...

Страница 74: ...ithout volume User entry Signed floating point number Factory setting 0 Spec enth fac Navigation Expert Sensor System units User spec units Spec enth fac 0583 Description Use this function to enter a...

Страница 75: ...cription Use this function to enter the offset for adapting the user specific energy unit without time User entry Signed floating point number Factory setting 0 Energy factor Navigation Expert Sensor...

Страница 76: ...specific pressure unit User entry Signed floating point number Factory setting 0 Pressure factor Navigation Expert Sensor System units User spec units Pressure factor 0579 Description Use this functio...

Страница 77: ...on time of the device also increases User entry 0 to 100 0 s Factory setting 0 s Additional information Description The damping is performed by a PT1 element 2 User entry Value 0 no damping Value 0 da...

Страница 78: ...f the temperature measured value User entry 0 to 999 9 s Factory setting 0 s Additional information Description The damping is performed by a PT1 element 4 User entry Value 0 no damping Value 0 dampin...

Страница 79: ...ation Expert Sensor Process param Low flow cut off Low flow cut off Assign variable 1837 79 On value 1805 79 Off value 1804 80 Pres shock sup 1806 81 Assign variable Navigation Expert Sensor Process p...

Страница 80: ...able parameter 79 Off value Navigation Expert Sensor Process param Low flow cut off Off value 1804 Prerequisite One of the following options is selected in the Assign variable parameter 79 Mass flow V...

Страница 81: ...n Pressure shock suppression is enabled Prerequisite Flow rate on value of low flow cut off or Changing the flow direction Output values Flow displayed 0 Totalizer the totalizers are pegged at the las...

Страница 82: ...low falls below the on value of the low flow cut off pressure shock suppression is activated 3 The time entered has elapsed pressure shock suppression is deactivated 4 The actual flow value is now dis...

Страница 83: ...escription Use this function to enter a lower limit value to enable detection of empty or partially filled measuring tubes If the measured density falls below this value monitoring is enabled User ent...

Страница 84: ...Expert Sensor Process param Partial pipe det Response time 1859 Prerequisite One of the following options is selected in the Assign variable parameter 83 Density Ref density Description Use this funct...

Страница 85: ...rmally the value of the oscillation damping is 500 If the pipe is partially filled the value of the oscillation damping is 5000 A practical damping value would then be 2000 enter 2000 as the value 3 2...

Страница 86: ...trog ox NOx Nitrogen N2 Nitrous ox N2O Methane CH4 Hydrogen H2 Helium He Hydrog chlor HCl Hydrog sulf H2S Ethylene C2H4 Carbon diox CO2 Carbon monox CO Chlorine Cl2 Butane C4H10 Propane C3H8 Propylene...

Страница 87: ...ion Use this function to enter a temperature coefficient for the sound velocity of the gas User entry Positive floating point number Factory setting 0 m s K 3 2 5 External comp submenu Navigation Expe...

Страница 88: ...cted in the Pressure compen parameter 87 Description Use this function to enter a value for the process pressure that is used for pressure correction User entry Positive floating point number Factory...

Страница 89: ...ernal value The temperature value measured internally temperature sensor of the measuring sensor is used for compensation Current input 1 option Current input 2 option The temperature value read in vi...

Страница 90: ...Expert Sensor Calculated value Corr vol flow Corr vol flow Corr vol flow 90 Ext ref density 91 Fix ref density 91 Ref temperature 92 Linear exp coeff 92 Square exp coeff 93 Corr vol flow Navigation Ex...

Страница 91: ...w and is issued as an output signal Ext ref density Navigation Expert Sensor Calculated value Corr vol flow Ext ref density 6198 Prerequisite In the Corr vol flow parameter 90 the Ext ref density opti...

Страница 92: ...er 67 Reference density calculation n 1 t t 2 A0023403 N reference density fluid density currently measured t fluid temperature currently measured tN reference temperature at which the reference densi...

Страница 93: ...erence density User entry Signed floating point number Factory setting 0 0 1 K 3 2 7 Sensor adjustment submenu Navigation Expert Sensor Sensor adjustm Sensor adjustm Install direct 1809 93 Zero point...

Страница 94: ...measuring device A healthy measuring device has a maximum zero point deviation of 100 compared to the factory setting of the measuring device calibration report Navigation Expert Sensor Sensor adjustm...

Страница 95: ...able adjust Mass flow offset 1831 95 Mass flow factor 1832 96 Vol flow offset 1841 96 Vol flow factor 1846 96 Density offset 1848 97 Density factor 1849 97 Corr vol offset 1866 97 Corr vol factor 1867...

Страница 96: ...dditional information Description Corrected value factor value offset Vol flow offset Navigation Expert Sensor Sensor adjustm Variable adjust Vol flow offset 1841 Description Use this function to ente...

Страница 97: ...ed value factor value offset Density factor Navigation Expert Sensor Sensor adjustm Variable adjust Density factor 1849 Description Use this function to enter a quantity factor for the density This mu...

Страница 98: ...alue factor value offset Ref dens offset Navigation Expert Sensor Sensor adjustm Variable adjust Ref dens offset 1868 Description Use this parameter to enter the zero point shift for the reference den...

Страница 99: ...try Signed floating point number Factory setting 0 K Additional information Description Corrected value factor value offset Temp factor Navigation Expert Sensor Sensor adjustm Variable adjust Temp fac...

Страница 100: ...calibration Zero point Navigation Expert Sensor Calibration Zero point 6195 Description Use this function to enter the zero point correction value for the sensor User entry Signed floating point numbe...

Страница 101: ...pert I O config I O config I O 1 to n terminals 3902 1 to n 101 I O 1 to n info 3906 1 to n 101 I O 1 to n type 3901 1 to n 102 Apply I O config 3907 102 Alteration code 2762 103 I O 1 to n terminals...

Страница 102: ...d for the fieldbus I O 1 to n type Navigation Expert I O config I O 1 to n type 3901 1 to n Prerequisite For the following order code Output input 2 option D Configurable I O initial setting off Outpu...

Страница 103: ...hange User entry Positive integer Factory setting 0 Additional information Description The I O configuration is changed in the I O type parameter 102 3 4 Input submenu Navigation Expert Input Input Cu...

Страница 104: ...not use any terminal numbers Signal mode Navigation Expert Input Current input 1 to n Signal mode 1610 1 to n Prerequisite The measuring device is not approved for use in the hazardous area with type...

Страница 105: ...0 Additional information Current input behavior The current input behaves differently depending on the settings configured in the following parameters Current span 104 Failure mode 106 Configuration e...

Страница 106: ...alue is used Defined value A user defined measured value is used Failure value parameter 106 Failure value Navigation Expert Input Current input 1 to n Failure value 1602 1 to n Prerequisite In the Fa...

Страница 107: ...vigation Expert Input Status input 1 to n Assign stat inp 1352 1 to n Description Use this function to select the function for the status input Selection Off Reset totaliz 1 Reset totaliz 2 Reset tota...

Страница 108: ...escription Use this function to determine the input signal level at which the assigned function is activated Selection High Low Factory setting High Response time Navigation Expert Input Status input...

Страница 109: ...1 to n 110 Current span 0353 1 to n 111 Fixed current 0365 1 to n 112 0 4 mA value 0367 1 to n 112 20 mA value 0372 1 to n 114 Measuring mode 0351 1 to n 115 Damping out 1 to n 0363 1 to n 120 Respons...

Страница 110: ...sign curr 1 to n Navigation Expert Output Curr output 1 to n Assign curr 1 to n 0359 1 to n Description Use this function to select a process variable for the current output Detailed description of th...

Страница 111: ...al on alarm Selection 4 20 mA NAMUR 4 20 mA US 4 20 mA 0 20 mA Fixed current Factory setting Country specific 4 20 mA NAMUR 4 20 mA US Additional information Description In the event of a device alarm...

Страница 112: ...gnostic message S441 Curr output 1 to n is displayed Fixed current Navigation Expert Output Curr output 1 to n Fixed current 0365 1 to n Prerequisite The Fixed current option is selected in the Curren...

Страница 113: ...ers Current span 111 Failure mode 122 Configuration examples Some examples of parameter settings and their effect on the current output are given in the following section Configuration example A Measu...

Страница 114: ...ith Rev flow comp option If flow is characterized by severe fluctuations e g when using reciprocating pumps flow components outside the measuring range are buffered balanced and output after a maximum...

Страница 115: ...s displayed Configuration examples Observe the configuration examples for the 0 4 mA value parameter 112 Measuring mode Navigation Expert Output Curr output 1 to n Measuring mode 0351 1 to n Prerequis...

Страница 116: ...ffective flow exceeds or falls below this measuring range the diagnostic message S441 Curr output 1 to n is displayed Forward Reverse option Q 20 0 I mA 4 3 1 2 0 A0013758 I Current Q Flow 1 Value ass...

Страница 117: ...es Example 1 Defined measuring range lower range value and upper range value with the same sign Q 20 4 0 I mA 1 2 A0028084 2 Measuring range I Current Q Flow 1 Lower range value value assigned to 0 4...

Страница 118: ...elay of 60 s t S A A S I mA A0028094 I Current t Time S Flow components saved A Balancing of saved flow components Example 2 Defined measuring range lower range value and upper range value with differ...

Страница 119: ...iagnostic message S441 Curr output 1 to n is displayed b The current output signal is proportional to the process variable assigned t I mA a b A0028100 I Current t Time With Forward Reverse option Thi...

Страница 120: ...meter 18 One of the following options is selected in the Current span parameter 111 4 20 mA NAMUR 4 20 mA US 4 20 mA 0 20 mA Description Use this function to enter a time constant for the reaction tim...

Страница 121: ...ons Oscil frequency Oscil amplitude Oscil damping and Signal asymmetry Value 1 display parameter 18 One of the following options is selected in the Current span parameter 111 4 20 mA NAMUR 4 20 mA US...

Страница 122: ...onic temp Osc freq 0 Osc ampl 0 Freq fluct 0 Osc damping 0 Osc damp fluct 0 Signal asymmetry Exc current 0 Detailed description of the options Oscil frequency Oscil amplitude Oscil damping and Signal...

Страница 123: ...re the device alarm occurred Actual value option The current output adopts the measured value on the basis of the current flow measurement the device alarm is ignored Defined value option The current...

Страница 124: ...ion Expert Output PFS output 1 to n PFS output 1 to n Terminal no 0492 1 to n 125 Signal mode 0490 1 to n 126 Operating mode 0469 1 to n 126 Assign pulse 1 to n 0460 1 to n 128 Value per pulse 0455 1...

Страница 125: ...145 Assign dir check 0484 1 to n 145 Assign status 0485 1 to n 146 Switch on delay 0467 1 to n 146 Switch off delay 0465 1 to n 146 Failure mode 0486 1 to n 147 Switch status 1 to n 0461 1 to n 147 I...

Страница 126: ...9 1 to n Description Use this function to select the operating mode of the output as a pulse frequency or switch output Selection Pulse Frequency Switch Factory setting Pulse Additional information Pu...

Страница 127: ...flow carrier mass flow density reference density concentration temperature carrier tube temperature electronic temperature vibration frequency frequency fluctuation oscillation amplitude oscillation...

Страница 128: ...rating mode parameter 126 Description Use this function to select the process variable for the pulse output Selection Off Mass flow Volume flow Correct vol flow Target mass flow Carrier mass fl Target...

Страница 129: ...resolution the higher the frequency of the pulse response Pulse width Navigation Expert Output PFS output 1 to n Pulse width 0452 1 to n Prerequisite In the Operating mode parameter 126 the Pulse opti...

Страница 130: ...Output PFS output 1 to n Measuring mode 0457 1 to n Prerequisite In the Operating mode parameter 126 the Pulse option is selected and in the Assign pulse parameter 128 one of the following options is...

Страница 131: ...perating mode parameter 126 the Pulse option is selected and one of the following options is selected in the Assign pulse parameter 128 Mass flow Volume flow Correct vol flow Target mass flow Carrier...

Страница 132: ...or the duration of the pulse NO contact and is safety oriented NC NO t t 0 1 0 1 A0028726 0 Non conductive 1 Conductive NC NC contact normally closed NO NO contact normally open The output behavior ca...

Страница 133: ...Concentration Temperature Carr pipe temp Electronic temp Osc freq 0 Osc ampl 0 Freq fluct 0 Osc damping 0 Osc damp fluct 0 Signal asymmetry Exc current 0 HBSI Pressure Factory setting Off Min freq val...

Страница 134: ...freq value 0454 1 to n Prerequisite In the Operating mode parameter 126 the Frequency option is selected and one of the following options is selected in the Assign freq parameter 132 Mass flow Volume...

Страница 135: ...the options Oscil frequency Oscil amplitude Oscil damping and Signal asymmetry Value 1 display parameter 18 Description Use this function to enter the measured value for the start value frequency User...

Страница 136: ...information Description Use this function to enter the maximum measured value at the maximum frequency The selected process variable is output as a proportional frequency Dependency The entry depends...

Страница 137: ...tailed description of the configuration examples see the Measuring mode parameter 115 Damping out 1 to n Navigation Expert Output PFS output 1 to n Damping out 1 to n 0477 1 to n Prerequisite In the O...

Страница 138: ...sponse time Navigation Expert Output PFS output 1 to n Response time 0491 1 to n Prerequisite In the Operating mode parameter 126 the Frequency option is selected and in the Assign freq parameter 132...

Страница 139: ...option is selected and one of the following options is selected in the Assign freq parameter 132 Mass flow Volume flow Correct vol flow Target mass flow Carrier mass fl Density Ref density Concentrati...

Страница 140: ...Actual value option is only recommended if it can be guaranteed that all possible alarm conditions will not affect the measurement quality Failure freq Navigation Expert Output PFS output 1 to n Fail...

Страница 141: ...the switch output Selection Off On Diag behavior Limit Fl direct check Status Factory setting Off Additional information Selection Off The switch output is permanently switched off open non conductive...

Страница 142: ...ection Alarm The switch output signals only diagnostic events in the alarm category Alarm or warning The switch output signals diagnostic events in the alarm and warning category Warning The switch ou...

Страница 143: ...formation Description Behavior of status output when Switch on value Switch off value Process variable Switch on value transistor is conductive Process variable Switch off value transistor is non cond...

Страница 144: ...value transistor is non conductive t t 1 2 3 A B A0026893 1 Switch on value Switch off value 2 Conductive 3 Non conductive A Process variable B Status output Switch on value Navigation Expert Output...

Страница 145: ...nction to enter the measured value for the switch off point User entry Signed floating point number Factory setting Country specific 0 kg h 0 lb min Additional information Description Use this functio...

Страница 146: ...conductive Otherwise the switch output is non conductive Switch on delay Navigation Expert Output PFS output 1 to n Switch on delay 0467 1 to n Prerequisite The Switch option is selected in the Opera...

Страница 147: ...t input value Open In the event of a device alarm the switch output s transistor is set to non conductive Closed In the event of a device alarm the switch output s transistor is set to conductive Swit...

Страница 148: ...0 1 A0026693 Yes option passive positive 0 1 A0026692 3 5 3 Relay output 1 to n submenu Navigation Expert Output Relay output 1 to n Relay output 1 to n Terminal no 149 Relay outp func 149 Assign dir...

Страница 149: ...minal numbers Relay outp func Navigation Expert Output Relay output 1 to n Relay outp func 0804 1 to n Description Use this function to select an output function for the relay output Selection Closed...

Страница 150: ...requisite In the Relay outp func parameter 149 the Fl direct check option is selected Description Use this function to select a process variable for monitoring the flow direction Selection Off Volume...

Страница 151: ...p func parameter 149 the Diag behavior option is selected Description Use this function to select the category of the diagnostic events that are displayed for the relay output Selection Alarm Alarm or...

Страница 152: ...cription Use this function to enter the measured value for the switch off point User entry Signed floating point number Factory setting Country specific 0 kg h 0 lb min Additional information Descript...

Страница 153: ...switch on value closed conductive When using a hysteresis Switch on value Switch off value Dependency The unit is dependent on the process variable selected in the Assign limit parameter 150 Switch o...

Страница 154: ...the relay output s transistor is set to conductive Switch status Navigation Expert Output Relay output 1 to n Switch status 0801 1 to n Description Displays the current status of the relay output User...

Страница 155: ...Measuring mode 0984 158 Failure mode 0985 158 Pulse output 0987 159 Invert outp sig 0993 159 Master term no Navigation Expert Output Double pulse out Master term no 0981 Description Displays the maste...

Страница 156: ...ection Passive Active Passive NAMUR Factory setting Passive Assign pulse 1 Navigation Expert Output Double pulse out Assign pulse 1 0982 1 Description Use this function to select a process variable fo...

Страница 157: ...ue the better the resolution the higher the frequency of the pulse response Pulse width Navigation Expert Output Double pulse out Pulse width 0986 Description Use this function to enter the duration o...

Страница 158: ...absolute value but a distinction is not made between positive and negative flow Reverse flow Negative flow is output positive flow is not output Rev flow comp The flow components outside the span are...

Страница 159: ...ement quality such that the quality can no longer be guaranteed The Actual value option is only recommended if it can be guaranteed that all possible alarm conditions will not affect the measurement q...

Страница 160: ...Modbus config 160 Modbus info 165 Modbus data map 166 Web server 167 WLAN settings 170 3 6 1 Modbus config submenu Navigation Expert Communication Modbus config Modbus config Bus address 7112 161 Baud...

Страница 161: ...ction to select a transmission rate Selection 1200 BAUD 2400 BAUD 4800 BAUD 9600 BAUD 19200 BAUD 38400 BAUD 57600 BAUD 115200 BAUD Factory setting 19200 BAUD Data trans mode Navigation Expert Communic...

Страница 162: ...the sequence in which the bytes are transmitted The transmission sequence must be coordinated with the Modbus master Selection 0 1 2 3 3 2 1 0 1 0 3 2 2 3 0 1 Factory setting 1 0 3 2 Additional inform...

Страница 163: ...FLOAT Sequence Options 1 2 3 4 1 0 3 2 Byte 1 MMMMMMMM Byte 0 MMMMMMMM Byte 3 SEEEEEEE Byte 2 EMMMMMMM 0 1 2 3 Byte 0 MMMMMMMM Byte 1 MMMMMMMM Byte 2 EMMMMMMM Byte 3 SEEEEEEE 2 3 0 1 Byte 2 EMMMMMMM B...

Страница 164: ...re mode 7116 Description Use this function to select the measured value output in the event of a diagnostic message via Modbus communication Selection NaN value Last valid value Factory setting NaN va...

Страница 165: ...estrict access to the measuring device via fieldbus Modbus protocol Selection Read write Read only Factory setting Read write Additional information Description If read and or write protection is enab...

Страница 166: ...nterface 4 digit hexadecimal number 3 6 3 Modbus data map submenu Navigation Expert Communication Modbus data map Modbus data map Scan list reg 0 to 15 7114 166 Scan list reg 0 to 15 Navigation Expert...

Страница 167: ...9 Default gateway 7210 169 Webserver funct 7222 169 Login page 7273 170 Webserv language Navigation Expert Communication Web server Webserv language 7221 Description Use this function to select the We...

Страница 168: ...igation Expert Communication Web server DHCP client 7212 Description Use this function to activate and deactivate the DHCP client functionality Selection Off On Factory setting Off Additional informat...

Страница 169: ...ticular octet Factory setting 0 0 0 0 Webserver funct Navigation Expert Communication Web server Webserver funct 7222 Description Use this function to switch the Web server on and off Selection Off On...

Страница 170: ...der Factory setting With header 3 6 5 WLAN settings submenu Navigation Expert Communication WLAN settings WLAN settings WLAN 2702 171 WLAN mode 2717 171 SSID name 2714 171 Network security 2705 172 Se...

Страница 171: ...tion Use this function to enable and disable the WLAN connection Selection Disable Enable Factory setting Enable WLAN mode Navigation Expert Communication WLAN settings WLAN mode 2717 Description Use...

Страница 172: ...Additional information Selection Unsecured Access the WLAN connection without identification WPA2 PSK Access the WLAN connection with a network key Sec identific Navigation Expert Communication WLAN...

Страница 173: ...tet Factory setting 192 168 1 212 WLAN MAC address Navigation Expert Communication WLAN settings WLAN MAC address 2703 Expert Communication WLAN settings WLAN MAC address 2703 Description Displays the...

Страница 174: ...Description Use this function to select which name is used for the SSID 11 Selection Device tag User defined Factory setting User defined Additional information Selection Device tag The device tag nam...

Страница 175: ...actory setting Select antenna Navigation Expert Communication WLAN settings Select antenna 2713 Description Use this function to select whether the external or internal antenna is used for reception S...

Страница 176: ...tings IP address DNS 2720 Expert Communication WLAN settings IP address DNS 2720 Description Use this function to enter the IP address of the domain name server Factory setting 192 168 1 212 3 7 Appli...

Страница 177: ...nu Navigation Expert Application Totalizer 1 to n Totalizer 1 to n Assign variable 0914 1 to n 177 Unit totalizer 1 to n 0915 1 to n 178 Operation mode 0908 1 to n 180 Control Tot 1 to n 0912 1 to n 1...

Страница 178: ...only Assign variable parameter 177 is still displayed in the Totalizer 1 to n submenu All other parameters in the submenu are hidden Unit totalizer 1 to n Navigation Expert Application Totalizer 1 to...

Страница 179: ...s liq bbl us beer Imperial units bbl imp beer or SI units Nl Nm Sl Sm US units Sft Sgal us Sbbl us liq Imperial units Sgal imp or Other units None Factory setting Country specific kg lb Additional inf...

Страница 180: ...he forward and reverse flow direction are totalized and balanced against one another Net flow is registered in the flow direction Forward total Only the flow in the forward flow direction is totalized...

Страница 181: ...process is restarted Hold Totalizing is stopped Preset value 1 to n Navigation Expert Application Totalizer 1 to n Preset value 1 to n 0913 1 to n Prerequisite One of the following options is selected...

Страница 182: ...izers and the outputs This is specified in separate parameters Selection Stop The totalizer is stopped in the event of a device alarm Actual value The totalizer continues to count based on the actual...

Страница 183: ...gation Expert Diagnostics Diagnostics Actual diagnos 0691 184 Prev diagnostics 0690 184 Time fr restart 0653 185 Operating time 0652 185 Diagnostic list 186 Event logbook 190 Device info 192 Mainboard...

Страница 184: ...amp and corrective measures referring to the cause of the diagnostic message can be accessed via the key Example For the display format F271 Main electronics Timestamp Navigation Expert Diagnostics Ti...

Страница 185: ...ccurred User interface Days d hours h minutes m and seconds s Additional information Display The diagnostic message can be viewed via the Prev diagnostics parameter 184 Example For the display format...

Страница 186: ...Expert Diagnostics Diagnostic list Diagnostics 1 0692 Description Displays the current diagnostics message with the highest priority User interface Symbol for diagnostic behavior diagnostic code and...

Страница 187: ...for diagnostic behavior diagnostic code and short message Additional information Display Via the local display the time stamp and corrective measures referring to the cause of the diagnostic message c...

Страница 188: ...key Examples For the display format F271 Main electronics F276 I O module Timestamp Navigation Expert Diagnostics Diagnostic list Timestamp Description Displays the operating time when the diagnostic...

Страница 189: ...ccurred User interface Days d hours h minutes m and seconds s Additional information Display The diagnostic message can be viewed via the Diagnostics 4 parameter 188 Example For the display format 24d...

Страница 190: ...he diagnostic message can be viewed via the Diagnostics 5 parameter 189 Example For the display format 24d12h13m00s 3 8 2 Event logbook submenu Navigation Expert Diagnostics Event logbook Event logboo...

Страница 191: ...Selection All Failure F Funct check C Out of spec S Mainten req M Information I Factory setting All Additional information Description The status signals are categorized in accordance with VDI VDE 265...

Страница 192: ...time when error occurred Additional information Description A maximum of 20 event messages are displayed in chronological order If the Extended HistoROM application package order option is enabled in...

Страница 193: ...rmation User interface 1 XXXXXXXXX A0029422 1 Position of the header text on the display The number of characters displayed depends on the characters used Serial number Navigation Expert Diagnostics D...

Страница 194: ...transmitter User interface Promass300 500 Order code Navigation Expert Diagnostics Device info Order code 0008 Description Displays the device order code User interface Character string composed of l...

Страница 195: ...der code can also be found on the nameplate of the sensor and transmitter in the Ext ord cd field Ext order cd 2 Navigation Expert Diagnostics Device info Ext order cd 2 0021 Description Displays the...

Страница 196: ...O1 Main elec I O1 Software rev 196 Build no softw 196 Bootloader rev 196 Software rev Navigation Expert Diagnostics Main elec I O1 Software rev 0072 Description Use this function to display the softw...

Страница 197: ...ert Diagnostics Sens electronic Software rev 0072 Description Use this function to display the software revision of the module User interface Positive integer Build no softw Navigation Expert Diagnost...

Страница 198: ...by the I O module User interface Not used 26 27 I O 1 24 25 I O 2 22 23 I O 3 Software rev Navigation Expert Diagnostics I O module 2 Software rev 0072 Expert Diagnostics I O module 3 Software rev 007...

Страница 199: ...igation Expert Diagnostics I O module 2 Software rev 0072 Description Use this function to display the software revision of the module User interface Positive integer Build no softw Navigation Expert...

Страница 200: ...I O 1 terminals 3902 1 Description Displays the terminal numbers used by the I O module User interface Not used 26 27 I O 1 24 25 I O 2 22 23 I O 3 Software rev Navigation Expert Diagnostics I O modu...

Страница 201: ...avigation Expert Diagnostics Display module Display module Software rev 0072 201 Build no softw 0079 201 Bootloader rev 0073 202 Software rev Navigation Expert Diagnostics Display module Software rev...

Страница 202: ...he software User interface Positive integer 3 8 10 Data logging submenu Navigation Expert Diagnostics Data logging Data logging Assign chan 1 203 Assign chan 2 204 Assign chan 3 204 Assign chan 4 205...

Страница 203: ...flow Volume flow Correct vol flow Target mass flow Carrier mass fl Target vol flow Carrier vol fl Targ corr vol fl Carr corr vol fl Density Ref density Ref dens altern GSV flow GSVa NSV flow NSVa S W...

Страница 204: ...he option selected is changed Assign chan 2 Navigation Expert Diagnostics Data logging Assign chan 2 0852 Prerequisite The Extended HistoROM application package is available The software options curre...

Страница 205: ...option overv parameter 45 Description Use this function to enter the logging interval Tlog for data logging User entry 0 1 to 3600 0 s Factory setting 1 0 s Additional information Description This def...

Страница 206: ...data The logging data is cleared The logging process starts from the beginning Data logging Navigation Expert Diagnostics Data logging Data logging 0860 Description Use this function to select the dat...

Страница 207: ...s selected Description Use this function to start and stop measured value logging Selection None Delete start Stop Factory setting None Additional information Selection None Initial measured value log...

Страница 208: ...logging parameter 206 the Not overwriting option is selected Description Displays the total logging duration User interface Positive floating point number Factory setting 0 s Displ channel 1 submenu N...

Страница 209: ...XXXXX 175 77 40 69 kg h 100s A0016357 9 Chart of a measured value trend x axis depending on the number of channels selected displays 250 to 1000 measured values of a process variable y axis displays t...

Страница 210: ...t Diagnostics Data logging Displ channel 3 Prerequisite A process variable is defined in the Assign chan 3 parameter Description See the Display channel 1 parameter 208 Displ channel 4 submenu Navigat...

Страница 211: ...ium temp 213 Carr pipe temp 213 Oscil frequency 214 Oscil amplitude 215 Oscil damping 216 Signal asymmetry 216 Reset min max Navigation Expert Diagnostics Min max val Reset min max 6151 Description Us...

Страница 212: ...ighest previously measured temperature value of the main electronics module User interface Signed floating point number Additional information Dependency The unit is taken from the Temperature unit pa...

Страница 213: ...ue User interface Signed floating point number Additional information Dependency The unit is taken from the Temperature unit parameter 67 Maximum value Navigation Expert Diagnostics Min max val Medium...

Страница 214: ...on Dependency The unit is taken from the Temperature unit parameter 67 Maximum value Navigation Expert Diagnostics Min max val Carr pipe temp Maximum value 6029 Prerequisite For the following order co...

Страница 215: ...ighest previously measured oscillation frequency User interface Signed floating point number Oscil amplitude submenu Navigation Expert Diagnostics Min max val Oscil amplitude Oscil amplitude Minimum v...

Страница 216: ...in max val Oscil damping Minimum value 6122 Description Displays the lowest previously measured oscillation damping User interface Signed floating point number Maximum value Navigation Expert Diagnost...

Страница 217: ...igation Expert Diagnostics Min max val Signal asymmetry Maximum value 6014 Description Displays the highest previously measured signal asymmetry User interface Signed floating point number 3 8 12 Hear...

Страница 218: ...inp 1 to n 1609 1 to n 221 Curr out 1 to n sim 0354 1 to n 222 Value curr out 1 to n 0355 1 to n 222 FreqOutputSim 1 to n 0472 1 to n 222 Freq value 1 to n 0473 1 to n 223 Puls outp sim 1 to n 0458 1...

Страница 219: ...Off Mass flow Volume flow Correct vol flow Target vol flow Carrier vol fl Targ corr vol fl Carr corr vol fl Density Ref density Ref dens altern GSV flow GSVa NSV flow NSVa S W volume flow Water cut O...

Страница 220: ...s submenu 59 Status inp sim Navigation Expert Diagnostics Simulation Status inp sim 1355 Description Use this function to switch simulation of the status input on and off The display alternates betwee...

Страница 221: ...of the Function check category C while simulation is in progress The desired simulation value is defined in the Value curr inp 1 to n parameter Selection Off On Factory setting Off Additional informat...

Страница 222: ...ice is in normal measuring mode or another process variable is being simulated On Current simulation is active Value curr out 1 to n Navigation Expert Diagnostics Simulation Value curr out 1 to n 0355...

Страница 223: ...o n Navigation Expert Diagnostics Simulation Freq value 1 to n 0473 1 to n Prerequisite In the FreqOutputSim 1 to n parameter the On option is selected Description Use this function to enter a frequen...

Страница 224: ...ion Pulse value 1 to n 0459 1 to n Prerequisite In the Puls outp sim 1 to n parameter the Down count val option is selected Description Use this function to enter a pulse value for the simulation In t...

Страница 225: ...or the simulation In this way users can verify the correct adjustment of the switch output and the correct function of downstream switching units Selection Open Closed Additional information Selection...

Страница 226: ...is function to select a relay value for the simulation In this way users can verify the correct adjustment of the relay output and the correct function of downstream switching units Selection Open Clo...

Страница 227: ...avigation Expert Diagnostics Simulation Pulse value 0989 Prerequisite In the Puls outp sim parameter 226 the Down count val option is selected Description Use this function to enter a pulse value for...

Страница 228: ...228 Selection Sensor Electronics Configuration Process Factory setting Process Diag event sim Navigation Expert Diagnostics Simulation Diag event sim 0737 Description Use this function to select a di...

Страница 229: ...factory settings applie to the following parameters 20 mA value full scale value of the current output 100 bar graph value 1 For detailed information about the full scale value for measuring devices...

Страница 230: ...2 0 1 4 0 45 6 1 4 2 US units Only valid for USA and Canada 4 2 1 System units Mass lb Mass flow lb min Volume gal us Volume flow gal min us Corrected volume Sft Corrected volume flow Sft min Density...

Страница 231: ...d information about the pulse value for measuring devices for custody transfer see the Special Documentation for the device 7 Nominal diameter in lb p 0 002 0 02 0 02 0 2 4 2 5 On value low flow cut o...

Страница 232: ...d Kilogram time unit t s t min t h t d Metric ton time unit Ref density kg Nm kg Nl g Scm kg Sm Kilogram gram standard volume unit Corrected volume Nl Nm Sm Normal liter normal cubic meter standard c...

Страница 233: ...luid ounce gallon kilogallon million gallon bbl us liq bbl us beer bbl us oil bbl us tank Barrel normal liquids barrel beer barrel petrochemicals barrel filling tanks Volume flow af s af min af h af d...

Страница 234: ...p Mgal imp Gallon mega gallon bbl imp beer bbl imp oil Barrel beer barrel petrochemicals Volume flow gal s imp gal min imp gal h imp gal d imp Gallon time unit Mgal s imp Mgal min imp Mgal h imp Mgal...

Страница 235: ...ttings User entry Input range for the parameter Page number information and cross reference to the standard parameter description NOTICE If non volatile device parameters are modified via the MODBUS R...

Страница 236: ...Expert operating menu The following table provides an overview of the menu structure of the expert operating menu and its parameters The page reference indicates where the associated description of th...

Страница 237: ...der 0097 252 Header text 0112 252 Separator 0101 252 Contrast display 0105 252 Backlight 0111 252 Config backup 252 Operating time 0652 252 Last backup 2757 252 Config managem 2758 252 Backup state 27...

Страница 238: ...s 257 Mass flow unit 0554 257 Mass unit 0574 257 Volume flow unit 0553 258 Volume unit 0563 259 Cor volflow unit 0558 259 Corr vol unit 0575 260 Density unit 0555 260 Ref dens unit 0556 260 Temperatur...

Страница 239: ...d velocity 6147 262 Temp coeff SV 6181 262 External comp 262 Pressure compen 6130 262 Pressure value 6059 262 External press 6209 262 Temperature mode 6184 263 External temp 6080 263 Calculated value...

Страница 240: ...t 1 to n 265 Terminal no 1611 1 to n 265 Signal mode 1610 1 to n 265 Current span 1605 1 to n 265 0 4 mA value 1606 1 to n 265 20 mA value 1607 1 to n 265 Failure mode 1601 1 to n 265 Failure value 16...

Страница 241: ...n 267 Failure mode 0364 1 to n 267 Failure current 0352 1 to n 267 Output curr 1 to n 0361 1 to n 267 Measur curr 1 to n 0366 1 to n 267 PFS output 1 to n 267 Terminal no 0492 1 to n 267 Signal mode 0...

Страница 242: ...270 Switch out funct 0481 1 to n 270 Assign diag beh 0482 1 to n 270 Assign limit 0483 1 to n 270 Switch on value 0466 1 to n 270 Switch off value 0464 1 to n 270 Assign dir check 0484 1 to n 270 Ass...

Страница 243: ...ay 0814 1 to n 272 Failure mode 0811 1 to n 272 Switch status 0801 1 to n 273 Act relay state 0816 1 to n 273 Double pulse out 273 Master term no 0981 273 Slave term no 0990 273 Signal mode 0991 273 A...

Страница 244: ...Fieldb writ acc 7156 274 Modbus info 274 Device ID 7153 274 Device revision 7154 274 Modbus data map 275 Scan list reg 0 to 15 7114 275 Web server 275 Webserv language 7221 275 MAC Address 7214 275 I...

Страница 245: ...ion 276 Reset all tot 2806 276 Totalizer 1 to n 277 Assign variable 0914 1 to n 277 Unit totalizer 1 to n 0915 1 to n 277 Operation mode 0908 1 to n 277 Control Tot 1 to n 0912 1 to n 278 Preset value...

Страница 246: ...1 279 Serial number 0009 279 Firmware version 0010 279 Device name 0020 279 Order code 0008 279 Ext order cd 1 0023 279 Ext order cd 2 0021 279 Ext order cd 3 0022 279 ENP version 0012 279 Mainboard m...

Страница 247: ...max val 280 Reset min max 6151 280 Electronic temp 280 Medium temp 280 Carr pipe temp 281 Oscil frequency 281 Oscil amplitude 281 Oscil damping 281 Signal asymmetry 281 Data logging 282 Assign chan 1...

Страница 248: ...1 to n 1609 1 to n 283 Curr out 1 to n sim 0354 1 to n 283 Value curr out 1 to n 0355 1 to n 284 FreqOutputSim 1 to n 0472 1 to n 284 Freq value 1 to n 0473 1 to n 284 Puls outp sim 1 to n 0458 1 to n...

Страница 249: ...0 Operator 1 Maintenance 13 Ent access code 0003 2177 Integer Read Write 0 to 9999 13 6 3 1 System submenu Display submenu Navigation Expert System Display Parameter Register Data type Access Selectio...

Страница 250: ...ct 0 68 Freq fluct 0 73 Concentration 74 Target mass flow 75 Carrier mass fl 78 Carrier vol fl 79 Carr corr vol fl 83 Oil density 84 Water density 85 Density average 86 GSV flow 87 GSVa 88 Oil mass fl...

Страница 251: ...5 Density average 86 GSV flow 87 GSVa 88 Oil mass flow 89 Water mass flow 90 NSV flow 91 NSVa 92 S W volume flow 93 Ref dens altern 94 Oil corr vol fl 95 Water corr v fl 96 Targ corr vol fl 97 Target...

Страница 252: ...able 1 Enable 28 Visibility depends on order options or device settings Config backup submenu Navigation Expert System Config backup Parameter Register Data type Access Selection User entry User inter...

Страница 253: ...Off 1 Logbook only 2 Warning 3 Alarm 35 Diagnostic no 442 0658 4919 Integer Read Write 0 Off 1 Logbook only 2 Warning 3 Alarm 36 Diagnostic no 443 0659 5000 Integer Read Write 0 Off 1 Logbook only 2...

Страница 254: ...arm 40 Diagnostic no 944 0732 2082 Integer Read Write 0 Off 1 Logbook only 2 Warning 3 Alarm 40 Diagnostic no 948 0744 5179 Integer Read Write 0 Off 1 Logbook only 2 Warning 3 Alarm 41 Administration...

Страница 255: ...Register Data type Access Selection User entry User interface Mass flow 1838 2007 to 2008 Float Read Signed floating point number 47 Volume flow 1847 2009 to 2010 Float Read Signed floating point num...

Страница 256: ...stat inp 1353 1 to n 1 2746 2 4699 3 4700 Integer Read 0 Low 1 High 55 Output values submenu Value curr out 1 to n submenu Navigation Expert Sensor Measured val Output values Value curr out 1 to n Par...

Страница 257: ...r 59 Double pulse out submenu Navigation Expert Sensor Measured val Output values Double pulse out Parameter Register Data type Access Selection User entry User interface Pulse output 0987 7041 to 704...

Страница 258: ...1 hl min 22 hl h 23 hl d 24 Ml s 25 Ml min 26 Ml h 27 Ml d 32 af s 33 af min 34 af h 35 af d 36 ft s 37 ft min 38 ft h 39 ft d 40 fl oz s us 41 fl oz min us 42 fl oz h us 43 fl oz d us 44 gal s us 45...

Страница 259: ...us 90 kgal h us 91 kgal d us Volume unit 0563 2104 Integer Read Write 0 cm 1 dm 2 m 3 ml 4 l 5 hl 6 Ml Mega 8 af 9 ft 10 fl oz us 11 gal us 12 Mgal us 13 bbl us liq 14 bbl us beer 15 bbl us oil 16 bbl...

Страница 260: ...G15 C 10 SG20 C 11 lb ft 12 lb gal us 13 lb bbl us liq 14 lb bbl us beer 15 lb bbl us oil 16 lb bbl us tank 17 lb gal imp 18 lb bbl imp beer 19 lb bbl imp oil 21 g m 22 g ml 23 API 24 SG60 F 65 Ref de...

Страница 261: ...al characters 73 Density factor 0572 Float Read Write Signed floating point number 73 Density offset 0571 Float Read Write Signed floating point number 73 Pressure text 0581 String Read Write Max 10 c...

Страница 262: ...te 0 Liquid 1 Gas 85 Select gas type 6074 5229 Integer Read Write 0 Air 1 Nitrogen N2 2 Argon Ar 3 Helium He 4 Carbon diox CO2 5 Oxygen O2 6 Methane CH4 7 Ammonia NH3 9 Hydrogen H2 10 Ethane C2H6 11 P...

Страница 263: ...f temperature 1816 5136 to 5137 Float Read Write 273 15 to 99999 C 92 Linear exp coeff 1817 5132 to 5133 Float Read Write Signed floating point number 92 Square exp coeff 1818 5134 to 5135 Float Read...

Страница 264: ...point number 99 Calibration submenu Navigation Expert Sensor Calibration Parameter Register Data type Access Selection User entry User interface Cal factor 6025 7513 to 7514 Float Read Signed floatin...

Страница 265: ...at Read Write Signed floating point number 105 Failure mode 1601 1 to n 1 6159 2 6160 3 6161 Integer Read Write 1 Last valid value 2 Alarm 6 Defined value 106 Failure value 1602 1 to n 1 6163 to 6164...

Страница 266: ...32 Exc current 0 39 Electronic temp 48 Osc freq 0 63 Osc damping 0 66 Pressure 67 Osc damp fluct 0 68 Freq fluct 0 73 Concentration 74 Target mass flow 75 Carrier mass fl 78 Carrier vol fl 79 Carr co...

Страница 267: ...1 2 5912 3 5913 Integer Read Write 0 Min 1 Max 4 Actual value 5 Last valid value 6 Defined value 122 Failure current 0352 1 to n 1 5979 to 5980 2 5981 to 5982 3 5983 to 5984 Float Read Write 0 to 22 5...

Страница 268: ...Oil corr vol fl 95 Water corr v fl 96 Targ corr vol fl 97 Target vol flow 99 Oil volume flow 102 Water vol flow 128 Value per pulse 0455 1 to n 1 3034 to 3035 2 3036 to 3037 3 4714 to 4715 Float Read...

Страница 269: ...ns altern 94 Oil corr vol fl 95 Water corr v fl 96 Targ corr vol fl 97 Target vol flow 99 Oil volume flow 101 Water cut 102 Water vol flow 132 Min freq value 0453 1 to n 1 3526 to 3527 2 3528 to 3529...

Страница 270: ...2 Assign limit 0483 1 to n 1 3184 2 3185 3 4722 Integer Read Write 1 Mass flow 2 Volume flow 3 Correct vol flow 4 Density 5 Ref density 7 Temperature 16 Totalizer 1 17 Totalizer 2 18 Totalizer 3 66 Pr...

Страница 271: ...d Write 0 Actual status 1 Open 6 Closed 147 Switch status 1 to n 0461 1 to n 1 2485 2 2486 3 9917 Integer Read 1 Open 6 Closed 147 Invert outp sig 0470 1 to n 1 2583 2 2584 3 9916 Integer Read Write 0...

Страница 272: ...rg corr vol fl 97 Target vol flow 99 Oil volume flow 100 Oscil damping 101 Water cut 102 Water vol flow 150 Assign diag beh 0806 1 to n 1 8245 2 8246 3 8247 Integer Read Write 0 Alarm 1 Warning 2 Alar...

Страница 273: ...d 2 24 25 I O 2 3 22 23 I O 3 155 Signal mode 0991 5949 Integer Read Write 0 Passive 2 Active 3 Passive NAMUR 156 Assign pulse 1 0982 1 5993 Integer Read Write 0 Off 1 Mass flow 2 Volume flow 3 Correc...

Страница 274: ...00 BAUD 2 4800 BAUD 3 9600 BAUD 4 19200 BAUD 5 38400 BAUD 6 57600 BAUD 7 115200 BAUD 161 Data trans mode 7115 4913 Integer Read Write 0 RTU 1 ASCII 161 Parity 7122 4914 Integer Read Write 0 Even 1 Odd...

Страница 275: ...Write 0 English 1 Deutsch 2 Fran ais 3 Espa ol 4 Italiano 5 Nederlands 8 Svenska 10 Bahasa Indonesia 11 Japanese 12 Portuguesa 13 Polski 14 Ru 15 e tina Czech 16 Chinese 18 T rk e 19 ti ng Vi t Vit 20...

Страница 276: ...206 Integer Read Write 0 Unsecured 1 WPA2 PSK 2 EAP PEAP MSCHAP2 3 EAP TLS 4 EAP PEAP NoAuth 172 WLAN passphrase 2706 8611 to 8626 String Read Write 8 to 32 digit character string comprising numbers l...

Страница 277: ...90 NSV flow 91 NSVa 92 S W volume flow 94 Oil corr vol fl 95 Water corr v fl 96 Targ corr vol fl 97 Target vol flow 99 Oil volume flow 102 Water vol flow 177 Unit totalizer 1 to n 0915 1 to n 1 4604...

Страница 278: ...ad Symbol for diagnostic behavior diagnostic code and short message 184 Prev diagnostics 0690 2734 Integer Read Symbol for diagnostic behavior diagnostic code and short message 184 Time fr restart 065...

Страница 279: ...ss300 500 194 Order code 0008 2058 to 2067 String Read Character string composed of letters numbers and certain punctuation marks e g 194 Ext order cd 1 0023 2212 to 2221 String Read Character string...

Страница 280: ...ccess Selection User entry User interface Software rev 0072 7039 Integer Read Positive integer 201 Min max val submenu Navigation Expert Diagnostics Min max val Parameter Register Data type Access Sel...

Страница 281: ...litude submenu Navigation Expert Diagnostics Min max val Oscil amplitude Parameter Register Data type Access Selection User entry User interface Minimum value 6010 2472 to 2473 Float Read Signed float...

Страница 282: ...Oil mass flow 89 Water mass flow 90 NSV flow 91 NSVa 92 S W volume flow 93 Ref dens altern 94 Oil corr vol fl 95 Water corr v fl 96 Targ corr vol fl 97 Target vol flow 99 Oil volume flow 101 Water cu...

Страница 283: ...ration 74 Target mass flow 75 Carrier mass fl 78 Carrier vol fl 79 Carr corr vol fl 83 Oil density 84 Water density 85 Density average 86 GSV flow 87 GSVa 88 Oil mass flow 89 Water mass flow 90 NSV fl...

Страница 284: ...2 6220 3 6221 Integer Read Write 0 to 65535 224 Switch sim 1 to n 0462 1 to n 1 6223 2 6224 3 6225 Integer Read Write 0 Off 1 On 224 Switch status 1 to n 0463 1 to n 1 6227 2 6228 3 6229 Integer Read...

Страница 285: ...ion check Parameter 145 150 Assign frequency output Parameter 132 Assign limit Parameter 142 150 Assign process variable Parameter 79 83 177 Assign pulse output 1 Parameter 156 Assign pulse output 1 t...

Страница 286: ...vior of diagnostic no 444 0740 36 Assign behavior of diagnostic no 543 0643 37 Assign behavior of diagnostic no 830 0800 37 Assign behavior of diagnostic no 831 0641 38 Assign behavior of diagnostic n...

Страница 287: ...s 5 0696 189 Direct access 0106 11 Display damping 0094 26 Display interval 0096 25 Display language 0104 15 ENP version 0012 195 Enter access code 0003 13 Entire logging duration 0861 208 Extended or...

Страница 288: ...ameter 2807 100 Off value low flow cutoff 1804 80 On value low flow cutoff 1805 79 Operating mode Pulse frequency switch output 1 to n 0469 1 to n 126 Operating time 0652 29 43 185 Operating time from...

Страница 289: ...1 89 Temperature offset 1870 99 Temperature unit 0557 67 Terminal number Current input 1 to n 1611 1 to n 104 Current output 1 to n 0379 1 to n 109 Pulse frequency switch output 1 to n 0492 1 to n 125...

Страница 290: ...Format display Parameter 15 Frequency output simulation 1 to n Parameter 222 Frequency value 1 to n Parameter 223 Function see Parameter G Gateway IP address Parameter 176 H Header Parameter 26 Heade...

Страница 291: ...6 Pulse output simulation 1 to n Parameter 223 Pulse value Parameter 227 Pulse value 1 to n Parameter 224 Pulse width Parameter 129 157 Pulse frequency switch output 1 to n Submenu 56 124 R Received s...

Страница 292: ...58 154 Switch status 1 to n Parameter 57 147 225 226 Switch off delay Parameter 146 152 Switch off value Parameter 145 152 Switch on delay Parameter 146 153 Switch on value Parameter 144 153 System S...

Страница 293: ...Value status input 1 to n Submenu 55 Volume flow Parameter 48 Volume flow factor Parameter 96 Volume flow offset Parameter 96 Volume flow unit Parameter 61 Volume unit Parameter 63 W Web server Submen...

Страница 294: ...www addresses endress com...

Отзывы: