7.26 Integrated Fiber Optic Exam Light—P7915AT
Chapter 7: Accessories
Page 7 - 56
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
CAUTION:
Ensure the power cord is not pulled too tightly. Tension on the power
cord connection could result in equipment damage.
24. Using the cable tie (F) provided, secure the power cord (X) to the light
mounting plate (B).
25. Using the box knife, cut the excess cable tie securing the fiber optic cable
connector (M) to the stretcher.
26. Connect the fiber optic cable connector (M) to the appropriate connection
(I) on the illuminator box (A) (see figure 7-28 on page 7-52)
27. Place the gooseneck assembly (H) into its storage slot (K).
28. Plug the fiber optic exam light into an appropriate power source.
29. Turn the power switch to the ON position, and ensure the fiber optic exam
light works properly.
Cleaning
Refer to the section titled “Cleaning and Care” on page 6-3 for proper cleaning
instructions.
WARNING:
Do not touch the lamp inside the illuminator box immediately after use.
Allow a minimum of 30 minutes for the lamp to cool before replacing it.
Failure to do so could result in personal injury and equipment damage.
CAUTION:
Remove the integrated fiber optic exam light assembly before spray
washing the stretcher. Failure to do so could result in equipment
damage.