4.3 Electric Stretcher Siderail
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 16
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
15. Using the ratchet and the T25 Torx® head bit, install the two screws (P)
and cable clamps (N) along the upper frame assembly.
16. Install the wire ties (M) that secure the siderail cable assembly (K) and the
CPR cable (L) along the upper frame assembly.
17. Install the snap wire cover (J) (see figure 4-4 on page 4-13).
18. Cycle the siderail from the stored to the up position several times. Plug the
stretcher into an appropriate source, and test for proper operation.