4.12 Electric Stretcher Nurse Control Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 35
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
4.12 Electric Stretcher Nurse Control Switch
Tools required:
T25 Torx®
1
head bit
Ratchet
Removal
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury and equipment damage.
1. Unplug the stretcher from its power source.
2. Using the ratchet and the T25 Torx® head bit, remove the two screws (B)
and cable clamp (A) that secure the nurse control unit (D) to the frame
(see figure 4-15 on page 4-35).
3. Record the wiring orientation at the defective switch. This will aid in the
installation.
Figure 4-15. Electric Stretcher Nurse Control Switch
1. Torx® is a registered trademark of Textron, Inc.
m115b103