7.15 Upright Chest Cassette Holder (Old Style)
Chapter 7: Accessories
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 7 - 27
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
7
7.15 Upright Chest Cassette Holder (Old Style)
The TranStar® Trauma/OB/GYN Stretcher Upright Chest Cassette Holder
attaches to the back section of the stretcher and allows for x-ray shots of the
patient’s chest/back. It will support a cassette with a maximum weight of 12 lb
(5 kg).
Installation
Tools required:
Ratchet
T25 Torx®
1
head bit
Screwdriver
1. Raise the head section to the high position.
2. Using the ratchet and T25 Torx® bit, install the screws (A) that secure the
slider bracket (B) to the head section weldment (C) (see figure 7-14 on
page 7-27).
Figure 7-14. Upright Chest Cassette Holder (Old Style)
1. Torx® is a registered trademark of Textron, Inc.
m115b129