4.23 Procedural/Electric/Trauma/Surgical/OB/GYN Stretcher Upper Frame
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 74
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
Figure 4-36. OB/GYN Stretcher Upper Frame
Replacement
WARNING:
Use two people for this procedure. Failure to do so could result in
personal injury and equipment damage.
CAUTION:
Be careful not to damage the base shroud when you install the upper
frame assembly.
1. Lift the upper frame, and put the round crosstube (C) of the upper frame
onto the hydraulic cylinders. Ensure that the frame is in position on both
hydraulic cylinders.
m115d192
A
B
C
D