4.28 OB/GYN Stretcher Foot Support Gas Spring
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 87
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
4.28 OB/GYN Stretcher Foot Support Gas Spring
Tools required:
Ratchet
T30 Torx®
1
head bit
Adjustable wrench
T25 Torx® head bit
Phillips head screwdriver
5/32" hex head bit
Screwdriver
Side cutters
Removal
1. Remove the mattress assembly from the stretcher.
WARNING:
Make sure the stretcher is in the full flat position, and make sure it is
secure. Failure to do so could result in personal injury or equipment
damage.
2. Put the stretcher in the full flat position.
3. Remove the foot section from the stretcher.
4. Remove the foot support bellow by performing the following (see figure 4-
45 on page 4-87):
Figure 4-45. Foot Support Bellows
1. Torx® is a registered trademark of Textron, Inc.
m115d207
F
F
J
I
H
G
E