2.12 Push Handle Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 52
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
2.12 Push Handle Malfunction
1. Another part of the stretcher interferes with the push handle assembly.
Yes
No
↓
→
Go to step 4.
2. Take appropriate actions to eliminate the interference.
This solves the problem.
Yes
No
↓
→
Go to step 4.
3. Go to “Final Actions” on page 2-32.
4. The push handle falls into position when the release handle is activated.
Yes
No
↓
→
Replace the push handle latch (refer to procedure 7.10). If this
solves the problem, go to “Final Actions” on page 2-32.
Otherwise, continue to step 5.
5. The push handle locks into position.
Yes
No
↓
→
Replace the push handles (refer to procedure 7.10). If this solves
the problem, go to “Final Actions” on page 2-32. Otherwise, call
Hill-Rom Technical Support at (800) 445-3720.